Neutron detection

Neutron detection is the effective detection of neutrons entering a well-positioned detector. There are two key aspects to effective neutron detection: hardware and software. Detection hardware refers to the kind of neutron detector used (the most common today is the scintillation detector) and to the electronics used in the detection setup. Further, the hardware setup also defines key experimental parameters, such as source-detector distance, solid angle and detector shielding. Detection software consists of analysis tools that perform tasks such as graphical analysis to measure the number and energies of neutrons striking the detector.

Basic physics

Signatures by which a neutron may be detected

Atomic and subatomic particles are detected by the signature they produce through interaction with their surroundings. The interactions result from the particles' fundamental characteristics.


n0

p+
+
e
+
ν
e
.
Although the
p+
and
e
produced by neutron decay are detectable, the decay rate is too low to serve as the basis for a practical detector system.

Classic neutron detection options

As a result of these properties, detection of neutrons fall into several major categories:[3]

Types of neutron detectors

Gas proportional detectors

Gas proportional detectors can be adapted to detect neutrons. While neutrons do not typically cause ionization, the addition of a nuclide with high neutron cross-section allows the detector to respond to neutrons. Nuclides commonly used for this purpose are helium-3, lithium-6, boron-10 and uranium-235. Since these materials are most likely to react with thermal neutrons (i.e., neutrons that have slowed to equilibrium with their surroundings), they are typically surrounded by moderating materials to reduce their energy and increase the likelihood of detection.

Further refinements are usually necessary to differentiate the neutron signal from the effects of other types of radiation. Since the energy of a thermal neutron is relatively low, charged particle reactions are discrete (i.e., essentially monoenergetic and lie within a narrow bandwidth of energies) while other reactions such as gamma reactions will span a broad energy range, it is possible to discriminate among the sources.

As a class, gas ionization detectors measure the number (count rate), and not the energy of neutrons.

3He gas-filled proportional detectors

An isotope of Helium, 3He provides for an effective neutron detector material because 3He reacts by absorbing thermal neutrons, producing a 1H and 3H ion. Its sensitivity to gamma rays is negligible, providing a very useful neutron detector. Unfortunately the supply of 3He is limited to production as a byproduct from the decay of tritium (which has a 12.3 year half-life); tritium is produced either as part of weapons programs as a booster for nuclear weapons or as a byproduct of reactor operation.

BF3 gas-filled proportional detectors

As elemental boron is not gaseous, neutron detectors containing boron may alternately use boron trifluoride (BF3) enriched to 96% boron-10 (natural boron is 20% 10B, 80% 11B).[5] It should be noted that boron trifluoride is highly toxic.

Boron lined proportional detectors

Alternately, boron-lined gas-filled proportional counters react similarly to BF3 gas-filled proportional detectors, with the exception that the walls are coated with 10B. In this design, since the reaction takes place on the surface, only one of the two particles will escape into the proportional counter.

Scintillation neutron detectors

Scintillation neutron detectors include liquid organic scintillators,[6] crystals,[7][8] plastics, glass[9] and scintillation fibers.[10]

Neutron-sensitive scintillating glass fiber detectors

Scintillating 6Li glass for neutron detection was first reported in the scientific literature in 1957[11] and key advances were made in the 1960s and 1970s.[12][13] Scintillating fiber was demonstrated by Atkinson M. et al. in 1987[14] and major advances were made in the late 1980s and early 1990s at Pacific Northwest National Laboratory where it was developed as a classified technology.[15][16][17][18][19] It was declassified in 1994 and first licensed by Oxford Instruments in 1997, followed by a transfer to Nucsafe in 1999.[20][21][22] The fiber and fiber detectors are now manufactured and sold commercially by Nucsafe, Inc.[23]

The scintillating glass fibers work by incorporating 6Li and Ce3+ into the glass bulk composition. The 6Li has a high cross-section for thermal neutron absorption through the 6Li(n,α) reaction. Neutron absorption produces a tritium ion, an alpha particle, and kinetic energy. The alpha particle and triton interact with the glass matrix to produce ionization, which transfers energy to Ce3+ ions and results in the emission of photons with wavelength 390 nm - 600 nm as the excited state Ce3+ ions return to the ground state. The event results in a flash of light of several thousand photons for each neutron absorbed. A portion of the scintillation light propagates through the glass fiber, which acts as a waveguide. The fibers ends are optically coupled to a pair of photomultiplier tubes (PMTs) to detect photon bursts. The detectors can be used to detect both neutrons and gamma rays, which are typically distinguished using pulse-height discrimination. Substantial effort and progress in reducing fiber detector sensitivity to gamma radiation has been made. Original detectors suffered from false neutrons in a 0.02 mR gamma field. Design, process, and algorithm improvements now enable operation in gamma fields up to 20 mR/h (60Co).

The scintillating fiber detectors have excellent sensitivity, they are rugged, and have fast timing (~60 ns) so that a large dynamic range in counting rates is possible. The detectors have the advantage that they can be formed into any desired shape, and can be made very large or very small for use in a variety of applications.[24] Further, they do not rely on 3He or any raw material that has limited availability, nor do they contain toxic or regulated materials. Their performance matches or exceeds that of 3He tubes for gross neutron counting due to the higher density of neutron absorbing species in the solid glass compared to high-pressure gaseous 3He.[24] Even though the thermal neutron cross section of 6Li is low compared to 3He (940 barns vs. 5330 barns), the atom density of 6Li in the fiber is fifty times greater, resulting in an advantage in effective capture density ratio of approximately 10:1.

LiCaAlF6

LiCaAlF6 is a neutron sensitive inorganic scintillator crystal which like neutron-sensitive scintillating glass fiber detectors makes use of neutron capture by 6Li. Unlike scintillating glass fiber detectors however the 6Li is part of the crystalline structure of the scintillator giving it a naturally high 6Li density. A doping agent is added to provide the crystal with its scintillating properties, two common doping agents are cesium and europium. Europium doped LiCaAlF6 has the advantage over other materials that the number of optical photons produced per neutron capture is around 30.000 which is 5 times higher than for example in neutron-sensitive scintillating glass fibers.[25] This property makes neutron photon discrimination easier. Due to its high 6Li density this material is suitable for producing light weight compact neutron detectors, as a result LiCaAlF6 has been used for neutron detection at high altitudes on balloon missions.[26] The long decay time of Europium doped LiCaAlF6 makes it less suitable for measurements in high radiation environments, the cesium doped variant has a shorter decay time but suffers from a lower light-yield.

Semiconductor neutron detectors

Semiconductors have been used for neutron detection.[27]

Neutron activation detectors

Activation samples may be placed in a neutron field to characterize the energy spectrum and intensity of the neutrons. Activation reactions that have differing energy thresholds can be used including 56Fe(n,p) 56Mn, 27Al(n,α)24Na, 93Nb(n,2n) 92mNb, & 28Si(n,p)28Al.[28]

Fast neutron detectors

Fast neutrons are often detected by first moderating (slowing) them to thermal energies. However, during that process the information on the original energy of the neutron, its direction of travel, and the time of emission is lost. For many applications, the detection of “fret” neutrons that retain this information is highly desirable.[29]

Typical fast neutron detectors are liquid scintillators,[30] 4-He based noble gas detectors [31] and plastic detectors. Fast neutron detectors differentiate themselves from one another by their 1.) capability of neutron/gamma discrimination (through pulse shape discrimination) and 2.) sensitivity. The capability to distinguish between neutrons and gammas is excellent in noble gas based 4-He detectors due to their low electron density and excellent pulse shape discrimination property.

Detection of fast neutrons poses a range of special problems. A directional fast-neutron detector has been developed using multiple proton recoils in separated planes of plastic scintillator material. The paths of the recoil nuclei created by neutron collision are recorded; determination of the energy and momentum of two recoil nuclei allow calculation of the direction of travel and energy of the neutron that underwent elastic scattering with them.[32]

Applications

Neutron detection is used for varying purposes. Each application has different requirements for the detection system.

Experimental neutron detection

Experiments that make use of this science include scattering experiments in which neutrons directed and then scattered from a sample are to be detected. Facilities include the ISIS neutron source at the Rutherford Appleton Laboratory, the Spallation Neutron Source at the Oak Ridge National Laboratory, and the Spallation Neutron Source (SINQ) at the Paul Scherrer Institute, in which the neutrons are produced by spallation reaction, and the traditional research reactor facilities in which neutrons are produced during fission of uranium isotopes. Noteworthy among the various neutron detection experiments is the trademark experiment of the European Muon Collaboration, first performed at CERN and now termed the "EMC experiment." The same experiment is performed today with more sophisticated equipment to obtain more definite results related to the original EMC effect.

Challenges in neutron detection in an experimental environment

Neutron detection in an experimental environment is not an easy science. The major challenges faced by modern-day neutron detection include background noise, high detection rates, neutron neutrality, and low neutron energies.

Background noise

The main components of background noise in neutron detection are high-energy photons, which aren't easily eliminated by physical barriers. The other sources of noise, such as alpha and beta particles, can be eliminated by various shielding materials, such as lead, plastic, thermo-coal, etc. Thus, photons cause major interference in neutron detection, since it is uncertain if neutrons or photons are being detected by the neutron detector. Both register similar energies after scattering into the detector from the target or ambient light, and are thus hard to distinguish. Coincidence detection can also be used to discriminate real neutron events from photons and other radiation.

High detection rates

If the detector lies in a region of high beam activity, it is hit continuously by neutrons and background noise at overwhelmingly high rates. This obfuscates collected data, since there is extreme overlap in measurement, and separate events are not easily distinguished from each other. Thus, part of the challenge lies in keeping detection rates as low as possible and in designing a detector that can keep up with the high rates to yield coherent data.

Neutrality of neutrons

Neutrons are neutral and thus do not respond to electric fields. This makes it hard to direct their course towards a detector to facilitate detection. Neutrons also do not ionize atoms except by direct collision, so gaseous ionization detectors are ineffective.

Varying behavior with energy

Detectors relying on neutron absorption are generally more sensitive to low-energy thermal neutrons, and are orders of magnitude less sensitive to high-energy neutrons. Scintillation detectors, on the other hand, have trouble registering the impacts of low-energy neutrons.

Experimental setup and method

Figure 1 shows the typical main components of the setup of a neutron detection unit. In principle, the diagram shows the setup as it would be in any modern particle physics lab, but the specifics describe the setup in Jefferson Lab (Newport News, Virginia).

In this setup, the incoming particles, comprising neutrons and photons, strike the neutron detector; this is typically a scintillation detector consisting of scintillating material, a waveguide, and a photomultiplier tube (PMT), and will be connected to a data acquisition (DAQ) system to register detection details.

The detection signal from the neutron detector is connected to the scaler unit, gated delay unit, trigger unit and the oscilloscope. The scaler unit is merely used to count the number of incoming particles or events. It does so by incrementing its tally of particles every time it detects a surge in the detector signal from the zero-point. There is very little dead time in this unit, implying that no matter how fast particles are coming in, it is very unlikely for this unit to fail to count an event (e.g. incoming particle). The low dead time is due to sophisticated electronics in this unit, which take little time to recover from the relatively easy task of registering a logical high every time an event occurs. The trigger unit coordinates all the electronics of the system and gives a logical high to these units when the whole setup is ready to record an event run.

The oscilloscope registers a current pulse with every event. The pulse is merely the ionization current in the detector caused by this event plotted against time. The total energy of the incident particle can be found by integrating this current pulse with respect to time to yield the total charge deposited at the end of the PMT. This integration is carried out in the analog-digital converter (ADC). The total deposited charge is a direct measure of the energy of the ionizing particle (neutron or photon) entering the neutron detector. This signal integration technique is an established method for measuring ionization in the detector in nuclear physics.[34] The ADC has a higher dead time than the oscilloscope, which has limited memory and needs to transfer events quickly to the ADC. Thus, the ADC samples out approximately one in every 30 events from the oscilloscope for analysis. Since the typical event rate is around 106 neutrons every second,[35] this sampling will still accumulate thousands of events every second.

Separating neutrons from photons

The ADC sends its data to a DAQ unit that sorts the data in presentable form for analysis. The key to further analysis lies in the difference between the shape of the photon ionization-current pulse and that of the neutron. The photon pulse is longer at the ends (or "tails") whereas the neutron pulse is well-centered.[35] This fact can be used to identify incoming neutrons and to count the total rate of incoming neutrons. The steps leading to this separation (those that are usually performed at leading national laboratories, Jefferson Lab specifically among them) are gated pulse extraction and plotting-the-difference.

Gated pulse extraction

Ionization current signals are all pulses with a local peak in between. Using a logical AND gate in continuous time (having a stream of "1" and "0" pulses as one input and the current signal as the other), the tail portion of every current pulse signal is extracted. This gated discrimination method is used on a regular basis on liquid scintillators.[36] The gated delay unit is precisely to this end, and makes a delayed copy of the original signal in such a way that its tail section is seen alongside its main section on the oscilloscope screen.

After extracting the tail, the usual current integration is carried out on both the tail section and the complete signal. This yields two ionization values for each event, which are stored in the event table in the DAQ system.

Plotting the difference

Figure 2: Expected plot of tail energy against energy in the complete pulse plotted for all event energies. Dots represent number densities of events.

In this step lies the crucial point of the analysis: the extracted ionization values are plotted. Specifically, the graph plots energy deposition in the tail against energy deposition in the entire signal for a range of neutron energies. Typically, for a given energy, there are many events with the same tail-energy value. In this case, plotted points are simply made denser with more overlapping dots on the two-dimensional plot, and can thus be used to eyeball the number of events corresponding to each energy-deposition. A considerable random fraction (1/30) of all events is plotted on the graph.

If the tail size extracted is a fixed proportion of the total pulse, then there will be two lines on the plot, having different slopes. The line with the greater slope will correspond to photon events and the line with the lesser slope to neutron events. This is precisely because the photon energy deposition current, plotted against time, leaves a longer "tail" than does the neutron deposition plot, giving the photon tail more proportion of the total energy than neutron tails.

The effectiveness of any detection analysis can be seen by its ability to accurately count and separate the number of neutrons and photons striking the detector. Also, the effectiveness of the second and third steps reveals whether event rates in the experiment are manageable. If clear plots can be obtained in the above steps, allowing for easy neutron-photon separation, the detection can be termed effective and the rates manageable. On the other hand, smudging and indistinguishability of data points will not allow for easy separation of events.

Rate control

Detection rates can be kept low in many ways. Sampling of events can be used to choose only a few events for analysis. If the rates are so high that one event cannot be distinguished from another, physical experimental parameters (shielding, detector-target distance, solid-angle, etc.) can be manipulated to give the lowest rates possible and thus distinguishable events.

Finer detection points

It is important here to observe precisely those variables that matter, since there may be false indicators along the way. For example, ionization currents might get periodic high surges, which do not imply high rates but just high energy depositions for stray events. These surges will be tabulated and viewed with cynicism if unjustifiable, especially since there is so much background noise in the setup.

One might ask how experimenters can be sure that every current pulse in the oscilloscope corresponds to exactly one event. This is true because the pulse lasts about 50 ns, allowing for a maximum of 2×107 events every second. This number is much higher than the actual typical rate, which is usually an order of magnitude less, as mentioned above.[35] This means that is it highly unlikely for there to be two particles generating one current pulse. The current pulses last 50 ns each, and start to register the next event after a gap from the previous event.

Although sometimes facilitated by higher incoming neutron energies, neutron detection is generally a difficult task, for all the reasons stated earlier. Thus, better scintillator design is also in the foreground and has been the topic of pursuit ever since the invention of scintillation detectors. Scintillation detectors were invented in 1903 by Crookes but were not very efficient until the PMT (photomultiplier tube) was developed by Curran and Baker in 1944.[34] The PMT gives a reliable and efficient method of detection since it multiplies the detection signal tenfold. Even so, scintillation design has room for improvement as do other options for neutron detection besides scintillation.

See also

References

  1. 1 2 Particle Data Group's Review of Particle Physics 2006
  2. Particle Data Group Summary Data Table on Baryons
  3. 1 2 3 4 Tsoulfanidis, Nicholas (1995). Measurement and Detection of Radiation (2nd ed.). Washington, D.C.: Taylor & Francis. pp. 467–501. ISBN 1-56032-317-5.
  4. Materials with a high hydrogen content such as water or plastic
  5. Boron Trifluoride (BF3) Neutron Detectors
  6. Yousuke, I.; Daiki, S.; Hirohiko, K.; Nobuhiro, S.; Kenji, I. (2000). "Deterioration of pulse-shape discrimination in liquid organic scintillator at high energies". Nuclear Science Symposium Conference Record. IEEE. 1: 6/219–6/221. ISBN 0-7803-6503-8. doi:10.1109/NSSMIC.2000.949173.
  7. Kawaguchi, N.; Yanagida, T.; Yokota, Y.; Watanabe, K.; Kamada, K.; Fukuda, K.; Suyama, T.; Yoshikawa, A. (2009). "Study of crystal growth and scintillation properties as a neutron detector of 2-inch diameter eu doped LiCaAlF6 single crystal". Nuclear Science Symposium Conference Record. IEEE: 1493–1495. ISBN 978-1-4244-3961-4. doi:10.1109/NSSMIC.2009.5402299.
  8. Example crystal scintillator based neutron monitor.
  9. Bollinger, L. M.; Thomas, G. E.; Ginther, R. J. (1962). "Neutron Detection With Glass Scintillators". Nuclear Instruments and Methods. 17: 97–116. Bibcode:1962NucIM..17...97B. doi:10.1016/0029-554X(62)90178-7.
  10. Miyanaga, N.; Ohba, N.; Fujimoto, K. (1997). "Fiber scintillator/streak camera detector for burn history measurement in inertial confinement fusion experiment". Review of Scientific Instruments. 68 (1): 621–623. Bibcode:1997RScI...68..621M. doi:10.1063/1.1147667.
  11. Egelstaff, P. A.; et al. (1957). "Glass Scintillators For Prompt Detection Of Intermediate Energy Neutrons". Nuclear Instruments and Methods. 1: 197–199. Bibcode:1957NucIn...1..197E. doi:10.1016/0369-643x(57)90042-7.
  12. Bollinger, L. M.; Thomas, G. E.; Ginther, R. J. (1962). "Neutron Detection With Glass Scintillators". Nuclear Instruments and Methods. 17: 97–116. Bibcode:1962NucIM..17...97B. doi:10.1016/0029-554X(62)90178-7.
  13. Spowart, A. R. (1976). "Neutron Scintillating Glasses .1. Activation By External Charged-Particles And Thermal-Neutrons". Nuclear Instruments and Methods. 135: 441–453. Bibcode:1976NucIM.135..441S. doi:10.1016/0029-554X(76)90057-4.
  14. Atkinson, M.; Fent J.; Fisher C.; et al. (1987). "Initial Tests Of A High-Resolution Scintillating Fiber (Scifi) Tracker". Nuclear Instruments and Methods in Physics Research A. 254: 500–514. doi:10.1016/0168-9002(87)90022-2.
  15. Bliss, M.; Brodzinski R. L.; Craig R. A.; Geelhood B. D.; Knopf M. A.; Miley H. S.; Perkins R. W.; Reeder P. L.; Sunberg D. S.; Warner R. A.; Wogman N. A. (1995). "Glass-fiber-based neutron detectors for high- and low-flux environments". Proc. SPIE. 2551: 108. doi:10.1117/12.218622.
  16. Abel, K. H.; Arthur R. J.; Bliss M.; Brite D. W.; et al. (1993). "Performance and Applications of Scintillating-Glass-Fiber Neutron Sensors". Proceedings of the SCIFI 93 Workshop on Scintillating Fiber Detectors: 463–472.
  17. Abel, K. H.; Arthur R. J.; Bliss M.; Brite D. W.; et al. (1994). "Scintillating Glass Fiber-Optic Neutron Sensors". Materials Research Society Symposium Proceedings. 348: 203–208.
  18. Bliss, M.; Craig R. A.; Reeder P. L. (1994). "The Physics and Structure-property Relationships of Scintillator Materials: Effect of Thermal History and Chemistry on the Light Output of Scintillating Glasses". Nuclear Instruments and Methods in Physics Research A. 342: 357–393. Bibcode:1994NIMPA.342..357B. doi:10.1016/0168-9002(94)90263-1.
  19. Bliss, M.; Craig R. A.; Reeder P. L.; Sunberg D. S.; Weber M. J. (1994). "Relationship Between Microstructure and Efficiency of Scintillating Glasses". Materials Research Society Symposium Proceedings. 348: 195–202.
  20. Seymour, R.; Crawford, T.; et al. (2001). "Portal, freight and vehicle monitor performance using scintillating glass fiber detectors for the detection of plutonium in the Illicit Trafficking Radiation Assessment Program". Journal of Radioanalytical and Nuclear Chemistry. 248: 699–705.
  21. Seymour, R. S.; Craig R. A.; Bliss M.; Richardson B.; Hull C. D.; Barnett D. S. (1998). "Performance of a neutron-sensitive scintillating glass-fiber panel for portal, freight and vehicle monitoring". Proc. SPIE. 3536: 148–155. doi:10.1117/12.339067.
  22. Seymour, R. S.; Richardson B.; Morichi M.; Bliss M.; Craig R. A.; Sunberg D. S. (2000). "Scintillating-glass-fiber neutron sensors, their application and performance for plutonium detection and monitoring" (PDF). Journal of Radioanalytical and Nuclear Chemistry. 243: 387–388.
  23. Nucsafe Inc. website
  24. 1 2 Van Ginhoven, R. M.; Kouzes R. T.; Stephens D. L. (2009). "Alternative Neutron Detector Technologies for Homeland Security PIET-43741-TM-840 PNNL-18471".
  25. Yanagida, T.; et al. (2011). "Europium and Sodium Codoped LiCaAlF6 Scintillator for Neutron Detection". Applied Physics Express. 4: 106401. Bibcode:2011APExp...4j6401Y. doi:10.1143/apex.4.106401.
  26. Kole, M.; et al. (2013). "A Balloon-borne Measurement of High Latitude Atmospheric Neutrons Using a LiCAF Neutron Detector". Nuclear Science Symposium Conference Record.
  27. Mireshghi, A.; Cho, G.; Drewery, J. S.; Hong, W. S.; Jing, T.; Lee, H.; Kaplan, S. N.; Perez-Mendez, V. (1994). "High efficiency neutron sensitive amorphous silicon pixel detectors". Nuclear Science. IEEE. 41 (4 , Part: 1–2): 915–921. Bibcode:1994ITNS...41..915M. doi:10.1109/23.322831.
  28. van Eijk, C. W. E.; de Haas, J. T. M.; Dorenbos, P.; Kramer, K. W.; Gudel, H. U. (2005). "Development of elpasolite and monoclinic thermal neutron scintillators". Nuclear Science Symposium Conference Record. IEEE. 1: 239–243. ISBN 0-7803-9221-3. doi:10.1109/NSSMIC.2005.1596245.
  29. Stromswold, D.C.; AJ Peurrung; RR Hansen; PL Reeder (1999). "Direct Fast-Neutron Detection. PNNL-13068, Pacific Northwest National Laboratory, Richland, WA.".
  30. Pozzi, S. A.; J. L. Dolan; E. C. Miller; M. Flaska; S. D. Clarke; A. Enqvist; P. Peerani; M. A. Smith-Nelson; E. Padovani; J. B. Czirr; L. B. Rees (2011). "Evaluation of New and Existing Organic Scintillators for Fast Neutron Detection". Proceedings of the Institute of Nuclear Materials Management 52nd Annual Meeting on CD-ROM, Palm Desert, California, USA. July 17 – 22.
  31. Lewis, J.M.; R. P. Kelley; D. Murer; K. A. Jordan (2014). "Fission signal detection using helium-4 gas fast neutron scintillation detectors". Appl. Phys. Lett. 105: 014102. Bibcode:2014ApPhL.105a4102L. doi:10.1063/1.4887366.
  32. Vanier, P. E.; Forman, L.; Dioszegi, I.; Salwen, C.; Ghosh, V. J. (2007). "Calibration and testing of a large-area fast-neutron directional detector". Nuclear Science Symposium Conference Record. IEEE: 179–184. ISBN 978-1-4244-0922-8. doi:10.1109/NSSMIC.2007.4436312.
  33. John F. Beacom & Mark R. Vagins (2004). "Antineutrino Spectroscopy with Large Water Čerenkov Detectors". Physical Review Letters. 93 (17): 171101. Bibcode:2004PhRvL..93q1101B. PMID 15525063. arXiv:hep-ph/0309300Freely accessible. doi:10.1103/PhysRevLett.93.171101.
  34. 1 2 Leo, W. R. (1994). Techniques for Nuclear and Particle Physics Experiments. Springer.
  35. 1 2 3 Cerny, J. C., Dolemal, Z., Ivanov, M. P., Kuzmin, E. P., Svejda, J., Wilhelm, I. (2003). "Study of neutron response and n–γ discrimination by charge comparison method for small liquid scintillation detector". Nuclear Instruments and Methods in Physics Research A. 527 (3): 512–518. Bibcode:2004NIMPA.527..512C. arXiv:nucl-ex/0311022Freely accessible. doi:10.1016/j.nima.2004.03.179.
  36. Jastaniah, S. D., Sellin, P. J. (2003). "Digital techniques for n–γ pulse shape discrimination capture-gated neutron spectroscopy using liquid". Nuclear Instruments and Methods in Physics Research A. 517: 202–210. Bibcode:2004NIMPA.517..202J. doi:10.1016/j.nima.2003.08.178.

Further reading

This article is issued from Wikipedia. The text is licensed under Creative Commons - Attribution - Sharealike. Additional terms may apply for the media files.