Scanning probe microscopy

Scanning probe microscopy (SPM) is a branch of microscopy that forms images of surfaces using a physical probe that scans the specimen. SPM was founded in 1981, with the invention of the scanning tunneling microscope, an instrument for imaging surfaces at the atomic level.

Many scanning probe microscopes can image several interactions simultaneously. The manner of using these interactions to obtain an image is generally called a mode.

The resolution varies somewhat from technique to technique, but some probe techniques reach a rather impressive atomic resolution. This is due largely because piezoelectric actuators can execute motions with a precision and accuracy at the atomic level or better on electronic command. This family of techniques can be called "piezoelectric techniques". The other common denominator is that the data are typically obtained as a two-dimensional grid of data points, visualized in false color as a computer image.

Established types of scanning probe microscopy

Of these techniques AFM and STM are the most commonly used for roughness measurements.

Image formation

To form images, scanning probe microscopes raster scan the tip over the surface. At discrete points in the raster scan a value is recorded (which value depends on the type of SPM and the mode of operation, see below). These recorded values are displayed as a heat map to produce the final STM images, usually using a black and white or an orange color scale.

Constant interaction mode

In constant interaction mode (often referred to as "in feedback"), a feedback loop is used to physically move the probe closer to or further from the surface (in the z axis) under study to maintain a constant interaction. This interaction depends on the type of SPM, for scanning tunneling microscopy the interaction is the tunnel current, for contact mode AFM or MFM it is the cantilever deflection, etc. The type of feedback loop used is usually a PI-loop, which is a PID-loop where the differential gain has been set to zero (as it amplifies noise). The z position of the tip (scanning plane is the xy-plane) is recorded periodically and displayed as a heat map. This is normally referred to as a topography image.

In this mode a second image, known as the ″error signal" or "error image" is also taken, which is a heat map of the interaction which was fed back on. Under perfect operation this image would be a blank at a constant value which was set on the feedback loop. Under real operation the image shows noise and often some indication of the surface structure. The user can use this image to edit the feedback gains to minimise features in the error signal.

If the gains are set incorrectly, many imaging artifacts are possible. If gains are too low features can appear smeared. If the gains are too high the feedback can become unstable and oscillate, producing striped features in the images which are not physical.

Constant height mode

In constant height mode the probe is not moved in the z-axis during the raster scan. Instead the value of the interaction under study is recorded (i.e. the tunnel current for STM, or the cantilever oscillation amplitude for amplitude modulated non-contact AFM). This recorded information is displayed as a heat map, and is usually referred to as a constant height image.

Constant height imaging is much more difficult than constant interaction imaging as the probe is much more likely to crash into the sample surface. Usually before performing constant height imaging one must image in constant interaction mode to check the surface has no large contaminants in the imaging region, to measure and correct for the sample tilt, and (especially for slow scans) to measure and correct for thermal drift of the sample. Piezoelectric creep can also be a problem, so the microscope often needs time to settle after large movements before constant height imaging can be performed.

Constant height imaging can be advantageous for eliminating the possibility of feedback artifacts.

Probe tips

The nature of an SPM probe depends entirely on the type of SPM being used. However, certain characteristics are common to all, or at least most, SPMs.

Most importantly the probe must have a very sharp apex. The apex of the probe defines the resolution of the microscope, the sharper the probe the better the resolution. For atomic resolution imaging the probe must be terminated by a single atom.

For many cantilever based SPMs (e.g. AFM and MFM), the entire cantilever and integrated probe are fabricated by acid [etching],[30] usually from silicon nitride. Conducting probes, needed for STM and SCM among others, are usually constructed from platinum/iridium wire for ambient operations, or tungsten for UHV operation. Other materials such as gold are sometimes used either for sample specific reasons or if the SPM is to be combined with other experiments such as TERS. Platinum/iridium (and other ambient) probes are normally cut using sharp wire cutters, the optimal method is to cut most of the way through the wire and then pull to snap the last of the wire, increasing the likelihood of a single atom termination. Tungsten wires are usually electrochemically etched, following this the oxide layer normally needs to be removed once the tip is in UHV conditions.

It is not uncommon for SPM probes (both purchased and "home-made") to not image with the desired resolution. This could be a tip which is too blunt or the probe may have more than one peak, resulting in a doubled or ghost image. For some probes, in situ modification of the tip apex is possible, this is usually done by either crashing the tip into the surface or by applying a large electric field. The latter is achieved by applying a bias voltage (of order 10V) between the tip and the sample, as this distance is usually 1-3 Angstroms, a very large field is generated.

Advantages of scanning probe microscopy

Disadvantages of scanning probe microscopy

Visualization and analysis software

In all instances and contrary to optical microscopes, rendering software is necessary to produce images. Such software is produced and embedded by instrument manufacturers but also available as an accessory from specialized work groups or companies. The main packages used are

References

  1. Binnig, G.; C. F. Qate; Ch. Gerber (1986-03-03). "Atomic Force Microscope". Physical Review Letters 56 (9): 930–933. Bibcode:1986PhRvL..56..930B. doi:10.1103/PhysRevLett.56.930. PMID 10033323.
  2. Kaiser, W. J.; L. D. Bell (1988). "Direct investigation of subsurface interface electronic structure by ballistic-electron-emission microscopy". Physical Review Letters 60 (14): 1406–1409. Bibcode:1988PhRvL..60.1406K. doi:10.1103/PhysRevLett.60.1406. PMID 10038030.
  3. Zhang, L.; T. Sakai, N. Sakuma, T. Ono, K. Nakayama; Sakuma, N.; Ono, T.; Nakayama, K. (1999). "Nanostructural conductivity and surface-potential study of low-field-emission carbon films with conductive scanning probe microscopy". Applied Physics Letters 75 (22): 3527–3529. Bibcode:1999ApPhL..75.3527Z. doi:10.1063/1.125377.
  4. Higgins, S. R.; R. J. Hamers (March 1996). Morphology and dissolution processes of metal sulfide minerals observed with the electrochemical scanning tunneling microscope. Journal of Vacuum Science and Technology B 14 (AVS). pp. 1360–1364. doi:10.1116/1.589098. Retrieved 2009-10-05.
  5. Weaver, J. M. R.; David W. Abraham (1991). "High resolution atomic force microscopy potentiometry". Journal of Vacuum Science and Technology B 9 (3): 1559–1561. Bibcode:1991JVSTB...9.1559W. doi:10.1116/1.585423.
  6. Meister, André; Gabi, Michael; Behr, Pascal; Studer, Philipp; Vörös, János; Niedermann, Philippe; Bitterli, Joanna; Polesel-Maris, Jérôme; Liley, Martha; Heinzelmann, Harry; Zambelli, Tomaso (2009). "FluidFM: Combining Atomic Force Microscopy and Nanofluidics in a Universal Liquid Delivery System for Single Cell Applications and Beyond". Nano Letters 9 (6): 2501–2507. Bibcode:2009NanoL...9.2501M. doi:10.1021/nl901384x. ISSN 1530-6984. PMID 19453133.
  7. FluidFM - Home - A unique new tool for single cell biology and beyond
  8. Single-Cell Analysis & Manipulation | FluidFM | Nanosurf
  9. Cytosurge
  10. Fritz, M.; M. Radmacher, N. Petersen, H. E. Gaub (May 1994). "Visualization and identification of intracellular structures by force modulation microscopy and drug induced degradation". The 1993 international conference on scanning tunneling microscopy. The 1993 international conference on scanning tunneling microscopy. Beijing, China: AVS. pp. 1526–1529. doi:10.1116/1.587278. Retrieved 2009-10-05.
  11. R. V. Lapshin (2011). "Feature-oriented scanning probe microscopy". In H. S. Nalwa. Encyclopedia of Nanoscience and Nanotechnology (PDF) 14. USA: American Scientific Publishers. pp. 105–115. ISBN 1-58883-163-9.
  12. Nonnenmacher, M.; M. P. O'Boyle; H. K. Wickramasinghe (1991). "Kelvin probe force microscopy". Applied Physics Letters 58 (25): 2921–2923. Bibcode:1991ApPhL..58.2921N. doi:10.1063/1.105227.
  13. Hartmann, U. (1988). "Magnetic force microscopy: Some remarks from the micromagnetic point of view". Journal of Applied Physics 64 (3): 1561–1564. Bibcode:1988JAP....64.1561H. doi:10.1063/1.341836.
  14. Sidles, J. A.; J. L. Garbini, K. J. Bruland, D. Rugar, O. Züger, S. Hoen, C. S. Yannoni (1995). "Magnetic resonance force microscopy". Reviews of Modern Physics 67 (1): 249. Bibcode:1995RvMP...67..249S. doi:10.1103/RevModPhys.67.249.
  15. BETZIG, E.; J. K. TRAUTMAN, T. D. HARRIS, J. S. WEINER, R. L. KOSTELAK (1991-03-22). "Breaking the Diffraction Barrier: Optical Microscopy on a Nanometric Scale". Science 251 (5000): 1468–1470. Bibcode:1991Sci...251.1468B. doi:10.1126/science.251.5000.1468. PMID 17779440. Retrieved 2009-10-05.
  16. Roelofs, A.; U. Bottger, R. Waser, F. Schlaphof, S. Trogisch, L. M. Eng (2000). "Differentiating 180° and 90° switching of ferroelectric domains with three-dimensional piezoresponse force microscopy". Applied Physics Letters 77 (21): 3444–3446. Bibcode:2000ApPhL..77.3444R. doi:10.1063/1.1328049.
  17. Reddick, R. C.; R. J. Warmack; T. L. Ferrell (1989-01-01). "New form of scanning optical microscopy". Physical Review B 39 (1): 767. Bibcode:1989PhRvB..39..767R. doi:10.1103/PhysRevB.39.767. Retrieved 2009-10-05.
  18. Matey, J. R.; J. Blanc (1985). "Scanning capacitance microscopy". Journal of Applied Physics 57 (5): 1437–1444. Bibcode:1985JAP....57.1437M. doi:10.1063/1.334506.
  19. Eriksson, M. A.; R. G. Beck, M. Topinka, J. A. Katine, R. M. Westervelt, K. L. Campman, A. C. Gossard (1996-07-29). "Cryogenic scanning probe characterization of semiconductor nanostructures". Applied Physics Letters 69 (5): 671–673. Bibcode:1996ApPhL..69..671E. doi:10.1063/1.117801. Retrieved 2009-10-05.
  20. Chang, A. M.; H. D. Hallen, L. Harriott, H. F. Hess, H. L. Kao, J. Kwo, R. E. Miller, R. Wolfe, J. van der Ziel, T. Y. Chang (1992). "Scanning Hall probe microscopy". Applied Physics Letters 61 (16): 1974–1976. Bibcode:1992ApPhL..61.1974C. doi:10.1063/1.108334.
  21. Hansma, PK; B Drake, O Marti, SA Gould, CB Prater (1989-02-03). "The scanning ion-conductance microscope". Science 243 (4891): 641–643. Bibcode:1989Sci...243..641H. doi:10.1126/science.2464851. PMID 2464851. Retrieved 2009-10-05.
  22. Wiesendanger, R.; M. Bode (2001-07-25). "Nano- and atomic-scale magnetism studied by spin-polarized scanning tunneling microscopy and spectroscopy". Solid State Communications 119 (4-5): 341–355. Bibcode:2001SSCom.119..341W. doi:10.1016/S0038-1098(01)00103-X. ISSN 0038-1098. Retrieved 2009-10-05.
  23. De Wolf, P.; J. Snauwaert, T. Clarysse, W. Vandervorst, L. Hellemans (1995). "Characterization of a point-contact on silicon using force microscopy-supported resistance measurements". Applied Physics Letters 66 (12): 1530–1532. Bibcode:1995ApPhL..66.1530D. doi:10.1063/1.113636.
  24. Xu, J. B.; K. Lauger, K. Dransfeld, I. H. Wilson (1994). "Thermal sensors for investigation of heat transfer in scanning probe microscopy". Review of Scientific Instruments 65 (7): 2262–2266. Bibcode:1994RScI...65.2262X. doi:10.1063/1.1145225.
  25. Binnig, G.; H. Rohrer, Ch. Gerber, E. Weibel (1982). "Tunneling through a controllable vacuum gap". Applied Physics Letters 40 (2): 178–180. Bibcode:1982ApPhL..40..178B. doi:10.1063/1.92999.
  26. Vorlesungsskript Physikalische Elektronik und Messtechnik (German)
  27. Trenkler, T.; P. De Wolf, W. Vandervorst, L. Hellemans (1998). "Nanopotentiometry: Local potential measurements in complementary metal--oxide--semiconductor transistors using atomic force microscopy". Journal of Vacuum Science and Technology B 16: 367–372. Bibcode:1998JVSTB..16..367T. doi:10.1116/1.589812.
  28. Volker Rose, John W. Freeland, Stephen K. Streiffer (2011). "New Capabilities at the Interface of X-Rays and Scanning Tunneling Microscopy". In Kalinin, Sergei V.; Gruverman, Alexei (Eds.). Scanning Probe Microscopy of Functional Materials: Nanoscale Imaging and Spectroscopy (1st ed.). New York: Springer. pp. 405–431. doi:10.1007/978-1-4419-7167-8_14. ISBN 978-1-4419-6567-7.
  29. Yoo, M. J.; Fulton, T. A. and Hess, H. F. and Willett, R. L. and Dunkleberger, L. N. and Chichester, R. J. and Pfeiffer, L. N. and West, K. W. (25 April 1997). "Scanning Single-Electron Transistor Microscopy: Imaging Individual Charges.". Science 276 (5312): 579–582. doi:10.1126/science.276.5312.579.
  30. Akamine, S.; Barrett, R. C.; Quate, C. F. (1990). "Improved atomic force microscope images using microcantilevers with sharp tips". Applied Physics Letters 57 (3): 316. Bibcode:1990ApPhL..57..316A. doi:10.1063/1.103677.
  31. Advanced Technologies Center - Advanced Technologies Center
  32. MountainsMap SPM

External links

Wikibooks has a book on the topic of: Nanowiki
This article is issued from Wikipedia - version of the Sunday, January 31, 2016. The text is available under the Creative Commons Attribution/Share Alike but additional terms may apply for the media files.