Pin Index Safety System

"PISS" redirects here. For other uses, see Piss (disambiguation).

The Pin Index Safety System, or PISS, is a safety system that uses geometric features on the yoke to ensure that pneumatic connections between a gas cylinder and a machine that uses pressurized gases are not connected to the wrong gas yoke. This system can be seen on an anesthesia machine.

Each gas cylinder has a pin configuration to fit its respective gas yoke. Refer to the figure on the right; units are in millimeters.

Standards

  • EN ISO 407 : Small medical gas cylinders - Pin-index yoke-type valve connections

The pin index system is a safety system (PISS) designed to ensure the correct gas is filled into the correct cylinder, and that the cylinder will only connect to the correct equipment. The positions of the holes on the cylinder valve correspond with the pins fitted to the yoke attached to the equipment. The pin positions for each medical gas are unique. If an attempt is made to fit the wrong gas cylinder to the yoke a tight seal will not be made, as the pins cannot locate.

References

    This article is issued from Wikipedia - version of the Wednesday, March 12, 2014. The text is available under the Creative Commons Attribution/Share Alike but additional terms may apply for the media files.