Microbeam
A microbeam is a narrow beam of radiation, of micrometer or sub-micrometer dimensions. Together with integrated imaging techniques, microbeams allow precisely defined quantities of damage to be introduced at precisely defined locations. Thus, the microbeam is a tool for investigators to study intra- and inter-cellular mechanisms of damage signal transduction.
A schematic of microbeam operation is shown on the right. Essentially, an automated imaging system locates user-specified targets, and these targets are sequentially irradiated, one by one, with a highly-focused radiation beam. Targets can be single cells, sub-cellular locations, or precise locations in 3D tissues. Key features of a microbeam are throughput, precision, and accuracy. While irradiating targeted regions, the system must guarantee that adjacent locations receive no energy deposition.
History
The first microbeam facilities were developed in the mid-90s. These facilities were a response to challenges in studying radiobiological processes using broadbeam exposures. Microbeams were originally designed to address two main issues:[1]
- The belief that the radiation-sensitivity of the nucleus was not uniform, and
- The need to be able to hit an individual cell with an exact number (particularly one) of particles for low dose risk assessment.
Additionally, microbeams were seen as ideal vehicles to investigate the mechanisms of radiation response.
Radiation-sensitivity of the cell
At the time it was believed that radiation damage to cells was entirely the result of damage to DNA. Charged particle microbeams could probe the radiation sensitivity of the nucleus, which at the time appeared not to be uniformly sensitive. Experiments performed at microbeam facilities have since shown the existence of a bystander effect. A bystander effect is any biological response to radiation in cells or tissues that did not experience a radiation traversal. These "bystander" cells are neighbors of cells that have experienced a traversal. The mechanism for the bystander effect is believed to be due to cell-to-cell communication. The exact nature of this communication is an area of active research for many groups.
Irradiation with an exact number of particles
At the low doses of relevance to environmental radiation exposure, individual cells only rarely experience traversals by an ionizing particle and almost never experience more than one traversal. For example, in the case of domestic radon exposure, cancer risk estimation involves epidemiological studies of uranium miners. These miners inhale radon gas, which then undergoes radioactive decay, emitting an alpha particle This alpha particle traverses the cells of the bronchial epithelium, potentially causing cancer. The average lifetime radon exposure of these miners is high enough that cancer risk estimates are driven by data on individuals whose target bronchial cells are subjected to multiple alpha particle traversals. On the other hand, for an average house occupant, about 1 in 2,500 target bronchial cells will be exposed per year to a single alpha particle, but less than 1 in 107 of these cells will experience traversals by more than one particle. Therefore, in order to extrapolate from miner to environmental exposures, it is necessary to be able to extrapolate from the effects of multiple traversals to the effects of single traversals of a particle.
Due to the random distribution of particle tracks, the biological effects of an exact number (particularly one) of particles cannot practically be simulated in the laboratory using conventional broadbeam exposures. Microbeam techniques can overcome this limitation by delivering an exact number (one or more) of particles per cell nucleus. True single-particle irradiations should allow measurement of the effects of exactly one alpha particle traversal, relative to multiple traversals. The application of such systems to low frequency processes such as oncogenic transformation depends very much on the technology involved. With an irradiation rate of at least 5,000 cells per hour, experiments with yields of the order of 10−4 can practically be accomplished. Hence, high throughput is a desired quality for microbeam systems.
Charged particle microbeam
The first microbeam facilities delivered charged particles. A charged particle microbeam facility must meet the following basic requirements:[2]
- The beam spot size should be on the order of a few micrometres or smaller, corresponding to cellular or sub-cellular dimensions.
- Irradiations of living cells should take place at atmospheric pressure.
- Beam current must be reduced to levels such that targets may be irradiated with an exact number of particles with high reproducibility.
- An imaging system is required to visualize and register cellular targets.
- Cell positioning must have high spatial resolution and reproducibility in order that the ion beam hit the target with a high degree of accuracy and precision.
- A particle detector with high efficiency must count the number of particles per target and switch off the beam after the desired number of particles have been delivered.
- Environmental conditions (humidity, for example) for cells must be maintained such that cells are under little or no stress.
Beam spot size
Beam spots with diameter down to about two micrometres can be obtained by collimating the beam with pinhole apertures or with a drawn capillary. Sub-micrometre beam spot sizes have been achieved by focusing the beam using various combinations of electrostatic or magnetic lenses. Both methods are used at present.
Vacuum window
A vacuum window is necessary in order to perform microbeam experiments on living cells. Generally, this is accomplished with the use of a vacuum-tight window of a polymer a few micrometres thick or 100-500 nm thick Silicon nitride.
Cell registration and positioning
Cells must be identified and targeted with a high degree of accuracy. This can be accomplished using cell staining and fluorescence microscopy or without staining through the use of techniques such as quantitative phase microscopy or phase contrast microscopy. Ultimately, the objective is to recognize cells, target them, and move them into position for irradiation as fast as possible. Throughputs of up to 15,000 cells per hour have been achieved.
Particle counters
Particles must be counted with a high degree of detection efficiency in order to guarantee that a specific number of ions are delivered to a single cell. Generally, detectors can be placed before or after the target to be irradiated. If the detector is placed after the target, the beam must have sufficient energy to traverse the target and reach the detector. If the detector is placed before the target, the detector must have a minimal effect on the beam. When the desired number of particles are detected, the beam is either deflected or shut off.
Other considerations
Living cells must be maintained under conditions that do not stress the cell, causing an unwanted biological response. Normally, cells must be attached to a substrate so that their position can be determined by the imaging system. Recent advancements in beam position control and high speed imaging have made flow through systems possible (Flow and Shoot).
X-ray microbeam
Some facilities have developed or are developing soft x-ray microbeams. In these systems, zone plates are used to focus characteristic x rays generated from a target hit by a charged particle beam. When using synchrotron x-rays as a source, x-ray microbeam can be obtained by cutting the beam with a precise slit system due to high directionality of synchrotron radiation.
Biological endpoint
Many biological endpoints have been studied including oncogenic transformation, apoptosis, mutations, and chromosomal aberrations.
Microbeam systems worldwide
Microbeam Facilities Worldwide[2] | Radiation Type/LET | Beam Spot Size on Cell | Running Biology? | |
---|---|---|---|---|
Radiological Research Accelerator Facility (RARAF),[3][4][5] Columbia University | any cation, x rays low to very high |
0.6 μm | yes | |
JAERI,[6][7][8] Takasaki, Japan | high |
yes | ||
Special Microbeam Utilization Research Facility (SMURF), Texas A&M | low |
no | ||
Superconducting Nanoscope for Applied nuclear (Kern-)physics Experiments (SNAKE),[9] University of Munich | From p to HI 2-10000 keV/μm |
0.5 μm | yes | |
INFN-LABEC,[10] Sesto Fiorentino, Florence, Italy | p, He, C other ions | 10 μm for 3 MeV p | no | |
INFN-LNL[11] Legnaro, Italy | p, 3He+,++,4He+,++ 7-150 keV/μm |
10 μm | yes | |
CENBG, Bordeaux, France | p, α Up to 3.5 MeV |
10 μm | ||
GSI,[12] Darmstadt, Germany | From α to U-ions Up to 11.4 MeV/n |
0.5 μm | yes | |
IFJ,[13] Cracow, Poland | p - Up to 2.5 MeV x ray - 4.5 keV |
12 μm 5 μm |
yes | |
LIPSION,[14] Leipzig, Germany | p, 4He+,++ Up to 3 MeV |
0.5 μm | yes | |
Lund NMP,[15] Lund, Sweden | p Up to 3 MeV |
5 μm | ||
CEA-LPS,[16] Saclay, France | p 4He+,++ Up to 3.75 MeV |
10 μm | yes | |
Queen's University, Belfast, Northern Ireland UK | x ray 0.3-4.5 keV |
< 1 μm | yes | |
University of Surrey, Guilford, UK | p, α, HI | 0.01 μm (in vacuum) | yes | |
PTB,[17] Braunschweig, Germany | p, α 3-200 keV/μm |
< 1 μm | yes | |
SPICE,[18][19] NIRS, Japan | p 3.4 MeV |
5 μm | yes | |
W-MAST, Tsuruga, Japan | p, He | 10 μm | no | |
McMaster University, Ontario, Canada | no | |||
Nagasaki University, Nagasaki, Japan | x-rays 0.3-4.5 keV |
< 1 μm | yes | |
Photon Factory,[20][21] KEK, Japan | x-rays 4-20 keV |
5 μm | yes | |
CAS-LIBB, Institute of Plasma Physics,[22][23] CAS, Hefei, China | p 2-3 MeV |
5 μm | yes | |
Centro Atómico Constituyentes, CNEA, Buenos Aires, Argentina | to H from U 15 MeV |
5 μm | yes | |
FUDAN University,[24] Shanghai, China | p,He 3 MeV |
2 μm | yes | |
Institute of Modern Physics[25] CAS, Lanzhou, China | ||||
Gray Laboratory, London | low, high | Yes | ||
Gray Laboratory, London | soft X | Yes | ||
PNL, Richland, Washington | low | Yes | ||
Padua, Italy | soft X | Yes | ||
MIT Boston | low, high | Yes | ||
L'Aquila, Italy | high | No | ||
LBL, Berkley | very high | No | ||
University of Maryland | low | Yes | ||
Tsukuba, Japan | soft X | Yes | ||
Nagatani, Japan | low, high | Yes | ||
Seoul, South Korea | low | Yes | ||
Helsinki, Finland | high | No | ||
Chapel Hill, North Carolina | low | No | ||
Gradignan, France | high | Yes | ||
Chiba, Japan |
Microbeam Workshops
There have been nine international workshops, held approximately once every two years, on Microbeam Probes of Cellular Radiation Response. These workshops serve as an opportunity for microbeam personnel to come together and share ideas. The proceedings of the workshops serve as an excellent reference on the state of microbeam-related science.
International Workshops on Microbeam Probes of Cellular Radiation Response | Year | Number of Microbeams |
---|---|---|
Gray Laboratory, London[1] | 1993 | 3 |
Pacific Northwest Labs, Washington | 1995 | 3 |
Columbia University, New York | 1997 | 4 |
Dublin, Ireland[26] | 1999 | 7 |
Stresa, Italy[27][28] | 2001 | 12 |
Oxford, England[29] | 2003 | 17 |
Columbia University, New York[30] | 2006 | 28 |
NIRS, Chiba, Japan[31] | 2008 | 31 |
GSI, Darmstadt, Germany | 2010 | |
Columbia University, New York | 2012 |
References
- 1 2 B.D. Michael, M. Folkard and K.M. Prise. Meeting Report: Microbeam Probes of Cellular Radiation Response, 4th L.H. Gray Workshop, 8–10 July 1993. Int. J. Radiat. Biol. 65:503-508 (1994). PMID 7908938
- 1 2 S. Gerardi. A comparative review of charged particle microbeam facilities. Radiation Protection Dosimetry 122:285-291 (2006). PMID 17132660
- ↑ Randers-Pehrson G, Geard CR, Johnson G, Elliston CD, Brenner DJ. The Columbia University single-ion microbeam. Radiat Res. 156:210-214 (2001). [PMID 11448243]
- ↑ Bigelow A., Ross G., Randers-Pehrson G. and Brenner DJ. The Columbia University Microbeam II endstation for cell imaging and irradiation. Nucl Instr Meth B 231:202-206 (2005).
- ↑ Bigelow A.W., Brenner D.J., Garty G. and Randers-Pehrson G. Single-particle/single-cell ion microbeams as probes of biological mechanisms. IEEE Trans. Plasma Sci. 36:1424-1431 (2008).
- ↑ Y. Kobayashi, T. Funayama, S. Wada, T. Sakashita. System of cell irradiation with a defined number of heavy ions Biol. Sci. Space. 16:105-106 (2002). PMID 12695571
- ↑ Y. Kobayashi, T. Funayama, S. Wada, T. Sakashita. System of cell irradiation with a precise number of heavy ions (II). Biol. Sci. Space. 17:253-254 (2003). PMID 14676403
- ↑ Y. Kobayashi, T. Funayama, S. Wada, T. Sakashita. System of cell irradiation with a defined number of heavy ions (III) Biol. Sci. Space. 18:186-187 (2004). PMID 15858384
- ↑ Hauptner A, Dietzel S, Drexler GA, Reichart P, Krücken R, Cremer T, Friedl AA, Dollinger G. Microirradiation of cells with energetic heavy ions. Radiat. Environ. Biophys. 42:237-245 (2004). PMID 14735370
- ↑ L. Giuntini, M. Massi, S. Calusi, The external scanning proton microprobe of Firenze: A comprehensive description, Nucl.Instr.&Meth.A 266-273 (2007), 576, Issue 2-3
- ↑ Gerardi S, Galeazzi G, Cherubini R. A microcollimated ion beam facility for |investigations of the effects of low-dose radiation. Radiat. Res. 164:586-590 (2005). PMID 16187793
- ↑ Heiss M, Fischer BE, Jakob B, Fournier C, Becker G, Taucher-Scholz G. Targeted irradiation of Mammalian cells using a heavy-ion microprobe. Radiat. Res. 165:231-239 (2006) PMID 16435921
- ↑ Veselov O, Polak W, Ugenskiene R, Lebed K, Lekki J, Stachura Z, Styczen J. Development of the IFJ single ion hit facility for cell irradiation.] Radiat. Prot. Dosimetry 122:316-319 (2006). PMID 17314088
- ↑ Anja Fiedler, Tilo Reinert, Judith Tanner, Tilman Butz. DNA double strand breaks and Hsp70 expression in proton irradiated living cells. Nucl Instr Meth B 260:169-173 (2007). http://dx.doi.org/10.1016/j.nimb.2007.02.020
- ↑ Pallon J, Malmqvist K. New applications of the nuclear microprobe for biological samples. Scanning Microsc. Suppl. 8:317-324 (1994). PMID 7638495
- ↑ Daudin L, Carrière M, Gouget B, Hoarau J, Khodja H. Development of a single ion hit facility at the Pierre Sue Laboratory: a collimated microbeam to study radiological effects on targeted living cells. Radiat. Prot. Dosimetry 122:310-312 (2006). PMID 17218368
- ↑ Greif K, Beverung W, Langner F, Frankenberg D, Gellhaus A, Banaz-Yasar F. The PTB microbeam: a versatile instrument for radiobiological research. Radiat Prot Dosimetry. 122: 313-315 (2006). PMID 17164277
- ↑ Hiroshi Yamaguchi, Yukio Sato, Hitoshi Imaseki, Nakahiro Yasuda, Tsuyoshi Hamano, Yoshiya Furusawa, Masao Suzuki, Takehiro Ishikawa, Teiji Mori, Kenichi Matsumoto, Teruaki Konishi, Masae Yukawa and Fuminori Soga. Single particle irradiation system to cell (SPICE) at NIRS. Nucl. Inst. Meth. B 210:292-295 (2003). http://dx.doi.org/10.1016/S0168-583X(03)01040-1
- ↑ Hitoshi Imaseki, Takahiro Ishikawa, Hiroyuki Iso, Teruaki Konishi, Noriyoshi Suya, Takeshi Hamano, Xufei Wang, Nakahiro Yasuda and Masae Yukawa. Progress report of the single particle irradiation system to cell (SPICE). Nucl. Inst. Meth. B 260:81-84 (2007). http://dx.doi.org/10.1016/j.nimb.2007.01.253
- ↑ K. Kobayashi, N. Usami, H. Maezawa, T. Hayashi, K. Hieda and K. Takakura. Synchrotron X-ray microbeam irradiation system for radiobiology. J. Biomed. Nanotech. 2:116-119 (2006).
- ↑ M. Maeda, N. Usami and K. Kobayashi. Low-dose Hypersensitivity in Nucleus-irradiated V79 Cells Studied with Synchrotron X-ray Microbeam. J. Radiat. Res. 49:171-180 (2008).
- ↑ X.F. Wang, X.H. Wang, L.Y. Chen, Z.W. Hu, J. Li, Y. Wu, B. Chen, S.H. Hu, J. Zhang, M.L. Xu, L. J. Wu, S.H. Wang, H.Y. Feng, F.R. Zhan, S.X. Peng, C.D. Hu, S.Q. Zhang, J.J. Chen, Z.T. Shi, H. Yuan, H.T. Yuan and Z.L. Yu. Development of the CAS- LIBB single-particle microbeam for localized irradiation of living cells, Chin.Sci. Bullet. 49:1806-1811(2004).http://www.springerlink.com/content/p9j2343951525t33/
- ↑ X.F.Wang, L.Y. Chen, Z.W. Hu, X.H. Wang, J.Zhang, J.Li, B.Chen,S.H.Hu, Z.T.Shi,Y.Wu, M.L.Xu, L.J.Wu, S.H.Wang,Z.L.Yu. Quantitative Single-Ion Irradiation by ASIPP Microbeam, Chin. Phys. Lett., 21: 821-824(2004).http://iopscience.iop.org/0256-307X/21/5/016/
- ↑ X.F.Wang, J.Q.Li, J.Z.Wang, J.X.Zhang, A.Liu, Z.J.He, W.Zhang, B.Zhang, C.L.Shao, L.Q.Shi. Current progress of the biological single-ion microbeam at FUDAN, Radiat Environ Biophys, DOI 10.1007/s00411-011-0361-1. PMID 21479813
- ↑ SHENG Lina, SONG Mingtao, ZHANG Xiaoqi,YANG Xiaotian, GAO Daqing, HE Yuan, ZHANG Bin,LIU Jie, SUN Youmei, DANG Bingrong,LI Wenjian, SU Hong, MAN Kaidi, GUO Yizhen, WANG Zhiguang, ZHAN Wenlong. Design of the IMP microbeam irradiation system for 100 MeV/u heavy ions. Chin. Phys. C.33: 315-320 (2009)
- ↑ Proceedings of the 4th International Workshop: Microbeam Probes of Cellular Radiation Response. Killiney Bay, Dublin, Ireland, July 17–18, 1999. Radiat. Res. 153:220–238 (2000)
- ↑ Proceedings of the 5th International Workshop: Microbeam Probes of Cellular Radiation Response Stresa, Lago Maggiore, Italy, May 26–27, 2001. Radiat. Res. 158:365-385 (2002).
- ↑ Brenner DJ, Hall EJ. Microbeams: a potent mix of physics and biology. Summary of the 5th International Workshop on Microbeam Probes of Cellular Radiation Response. Radiat. Prot. Dosimetry 99:283-286 (2002). PMID 12194307
- ↑ Proceedings of the 6th International Workshop/12th L. H. Gray Workshop: Microbeam Probes of Cellular Radiation Response St. Catherine's College, Oxford, United Kingdom, March 29–31, 2003. Radiat. Res. 161:87-119 (2004).
- ↑ Proceedings of the 7th International Workshop: Microbeam Probes of Cellular Radiation Response. Columbia University, New York, New York, March 15–17, 2006. Radiat. Res. 166:652-689 (2006).
- ↑ 8th International Workshop on Microbeam Probes of Cellular Radiation Response NIRS, Chiba, Japan, November 13–15, 2008. J. Radiat. Res. 50 Suppl.:A81-A125 (2009).