Thermal mass flow meter

There are three types of thermal mass flow meters in industry. Industrial thermal mass flow meters, also known as thermal dispersion or immersible mass flow meters comprise a family of instruments for the measurement of the total mass flow rate of a fluid, primarily gases, flowing through closed conduits. A second type is the capillary-tube type of thermal mass flow meter. Many mass flow controllers (MFC) which combine a mass flow meter, electronics and a valve are based on this design.

Both types measure fluid mass flow rate by means of the heat convected from a heated surface to the flowing fluid. In the case of the thermal dispersion, or immersible, type of flow meter, the heat is transferred to the boundary layer of the fluid flowing over the heated surface. In the case of the capillary-tube type, the heat is transferred to the bulk of the fluid flowing through a small heated capillary tube. The principles of operation of the two types are both thermal in nature, but are so substantially different that two separate standards are required. Additionally, their applications are much different. Thermal dispersion flow meters are commonly used for general industrial gas flow applications in pipes and ducts, whereas capillary types are primarily used for smaller flows of clean gases or liquids in tubes. This type is most widely used for thermal mass flow meters in industry. Nevertheless, the capillary type is not the subject of this discussion.

Thermal dispersion mass flow meter history

The operation of thermal dispersion mass flow meters is attributed to L.V. King who, in 1914, published his famous King's Law revealing how a heated wire immersed in a fluid flow measures the mass velocity at a point in the flow. King called his instrument a "hot-wire anemometer". However, it was not until the 1960s and 1970s that industrial-grade thermal dispersion mass flow meters finally emerged.

In operation

Temperature at the sensors varies depending upon the mass flow

Thermal mass flow meters are used almost entirely for gas flow applications. As the name implies, thermal mass flow meters use heat to measure flow. Thermal mass flow meters introduce heat into the flow stream and measure how much heat dissipates using one or more temperature sensors. The amount of heat lost from the sensor is dependent upon the sensor design and the thermal properties of the fluid. The thermal properties of the fluid can (and do) vary with pressure and temperature, however these variations are typically small in most applications. In these applications where the thermal properties of the fluid are known and relatively constant during actual operation, thermal flow meters can be used to measure the mass flow of the fluid because the thermal flow measurement is not dependent upon the pressure or temperature of the fluid.

This method works best with gas mass flow measurement because it is difficult to get a strong signal using thermal mass flow meters in liquids, due to considerations relating to heat absorption.

While all thermal flow meters use heat to make their flow measurements, there are two different methods for measuring how much heat is dissipated.

Constant temperature differential

One method is called the constant temperature differential method. Thermal flow meters using this method have two temperature sensors — a heated sensor and another sensor that measures the temperature of the gas. Mass flow rate is computed based on the amount of electrical power required to maintain a constant difference in temperature between the two temperature sensors.

Constant current

A second, and more popular concept, is called the constant current method. Thermal mass flow meters using this method also have a heated sensor and another one that senses the temperature of the flow stream. The power to the heated sensor is kept constant. Mass flow is measured as a function of the difference between the temperature of the heated sensor and the temperature of the flow stream. Both methods are based on the principle that higher velocity flows result in a greater cooling effect. Both measure mass flow based on the measured effects of cooling in the flow stream.

Industrial applications

The primary reason thermal mass flow meters are popular in industrial applications is the way they are designed and built. They feature no moving parts, nearly unobstructed straight through flow path, require no temperature or pressure corrections and retain accuracy over a wide range of flow rates. Straight pipe runs can be reduced by using dual-plate flow conditioning elements and installation is very simple with minimal pipe intrusions.

However, in many applications, the thermal properties of the fluid can be dependent upon fluid composition. In such applications, varying composition of the fluid during actual operation can affect the thermal flow measurement. Therefore, it is important for the thermal flow meter supplier to know the composition of the fluid so that the proper calibration factor can be used to determine the flow rate accurately. Suppliers can provide appropriate calibration information for other gas mixtures, however the accuracy of the thermal flow meter is dependent on the actual gas mixture being the same as the gas mixture used for calibration purposes. In other words, the accuracy of a thermal flow meter calibrated for a given gas mixture will be degraded if the actual flowing gas has a different composition.[1]