Peristaltic pump

From Wikipedia, the free encyclopedia
Peristaltic tube pump with two sprung rollers
Peristaltic pump in motion
Linear peristaltic pump

A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made). A rotor with a number of "rollers", "shoes", "wipers", or "lobes" attached to the external circumference of the rotor compresses the flexible tube. As the rotor turns, the part of the tube under compression is pinched closed (or "occludes") thus forcing the fluid to be pumped to move through the tube. Additionally, as the tube opens to its natural state after the passing of the cam ("restitution" or "resilience") fluid flow is induced to the pump. This process is called peristalsis and is used in many biological systems such as the gastrointestinal tract. Typically, there will be two or more rollers, or wipers, occluding the tube, trapping between them a body of fluid. The body of fluid is then transported, at ambient pressure, toward the pump outlet. Peristaltic pumps may run continuously, or they may be indexed through partial revolutions to deliver smaller amounts of fluid.

History

The peristaltic pump was first patented in the United States by Eugene Allen in 1881. It was popularized by heart surgeon Dr. Michael DeBakey while he was a medical student in 1932.[1]

Applications

Peristaltic pumps are typically used to pump clean/sterile or aggressive fluids because cross contamination with exposed pump components cannot occur. Some common applications include pumping IV fluids through an infusion device, aggressive chemicals, high solids slurries and other materials where isolation of the product from the environment, and the environment from the product, are critical. It is also used in heart-lung machines to circulate blood during a bypass surgery as the pump does not cause significant hemolysis.

Peristaltic pumps are also used in a wide variety of industrial applications. Their unique design makes them especially suited to pumping abrasives[2] and viscous fluids.[3]

Key design parameters

The peristaltic pump head and the rollers are of large diameter. The asymmetric head design and the spring-loaded off-center levers move the rollers gradually and softly, thus increasing tubing life and reducing pulsations.

The ideal peristaltic pump should have an infinite diameter of the pump head and the largest possible diameter of the rollers. Such an ideal peristaltic pump would offer an unlimited tubing lifetime and provide a constant and pulsation-free flow rate.[4]

Such an ideal peristaltic pump can not be constructed in reality. However, peristaltic pumps can be designed to approach these ideal peristaltic pump parameters. One example of a possible construction is depicted.

Chemical compatibility

The pumped fluid contacts only the inside surface of the tubing thereby negating concern for other valves, O-rings or seals that might be incompatible with fluid being pumped. Therefore, only the composition of the tubing that the pumped medium travels through is considered for chemical compatibility.

The tubing needs to be elastomeric to maintain the circular cross section after millions of cycles of squeezing in the pump. This requirement eliminates a variety of non-elastomeric polymers that have compatibility with a wide range of chemicals, such as PTFE, polyolefins, PVDF, etc. from consideration as material for pump tubing. The popular elastomers for pump tubing are nitrile (NBR), Hypalon, Viton, silicone, PVC, EPDM, EPDM+polypropylene (as in Santoprene), polyurethane and natural rubber. Of these materials, natural rubber has the best fatigue resistance, and EPDM and Hypalon have the best chemical compatibility. Silicone is popular with water-based fluids, such as in bio-pharma industry, but have limited range of chemical compatibility in other industries.

Extruded fluoropolymer tubes such as FKM (Viton, Fluorel, etc.) have good compatibility with acids, hydrocarbons, and petroleum fuels, but have insufficient fatigue resistance to achieve an effective tube life.

There are a couple of newer tubing developments that offer a broad chemical compatibility using lined tubing and fluoroelastomers.

With lined tubing, the thin inside liner is made of a chemically resistant material such as poly-olefin and PTFE that form a barrier for the rest of the tubing wall from coming in contact with the pumped fluid. These liners are materials that are often not elastomeric, therefore the entire tube wall cannot be made with this material for peristaltic pump applications. This tubing provides adequate chemical compatibility and life to be used in chemically challenging applications. There are a few things to keep in mind when using these tubes - any pin holes in the liner during manufacturing could render the tubing vulnerable to chemical attack. In the case of stiff plastic liners like the polyolefins, with repeated flexing in the peristaltic pump they can develop cracks, rendering the bulk material again vulnerable to chemical attack. A common issue with all lined tubing is delamination of the liner with repeated flexing that signals the end of the tube's life. For those with need for chemically compatible tubing, these lined tubing offer a good solution.

With fluoroelastomer tubing, the elastomer itself has the chemical resistance. In the case of e.g. Chem-Sure, it is made of a perfluoroelastomer, that has the broadest chemical compatibility of all elastomers. The two fluoroelastomer tubes listed above combine the chemical compatibility with a very long tube life stemming from their reinforcement technology, but come at a pretty high initial cost. One has to justify the cost with the total value derived over the long tube life, and compare with other options such as other tubing or even other pump technologies.

There are many online sites for checking the chemical compatibility of the tubing material with the pumped fluid. The manufacturers of these tubing may also have compatibility charts specific to their tubing.

While these charts cover a list of commonly encountered fluids, they may not have all the fluids. If there is a fluid whose compatibility is not listed anywhere, then a common test of compatibility is the immersion testing. A 1 to 2 inch sample of the tubing is immersed in the fluid to be pumped for anywhere from 24 to 48 hours, and the amount of weight change from before and after the immersion is measured. If the weight change is greater than 10% of the initial weight, then that tube is not compatible with the fluid, and should not be used in that application. This test is still a one way test, in the sense that there is still a remote chance that the tubing that passes this test can still be incompatible for the application since the combination of borderline compatibility and mechanical flexing can push the tube over the edge, resulting in premature tube failure.

In general, recent tubing developments have brought broad chemical compatibility to the peristaltic pump option that many chemical dosing applications can benefit over other current pump technologies.

Occlusion

The minimum gap between the roller and the housing determines the maximum squeeze applied on the tubing. The amount of squeeze applied to the tubing affects pumping performance and the tube life - more squeezing decreases the tubing life dramatically, while less squeezing can cause the pumped medium to slip back, especially in high pressure pumping, and decreases the efficiency of the pump dramatically and the high velocity of the slip back typically causes premature failure of the hose. Therefore, this amount of squeeze becomes an important design parameter.

The term "occlusion" is used to measure the amount of squeeze. It is either expressed as a percentage of twice the wall thickness, or as an absolute amount of the wall that is squeezed.

Let

y = occlusion
g = minimum gap between the roller and the housing
t = wall thickness of the tubing

Then

y = 2t - g (when expressed as the absolute amount of squeeze)
y = (2t - g) / (2t) × 100 (when expressed as a percentage of twice the wall thickness)

The occlusion is typically 10 to 20%, with a higher occlusion for a softer tube material and a lower occlusion for a harder tube material.

Thus for a given pump, the most critical tubing dimension becomes the wall thickness. An interesting point here is that the inside diameter of the tubing is not an important design parameter for the suitability of the tubing for the pump. Therefore, it is common for more than one ID be used with a pump, as long as the wall thickness remains the same.

Inside diameter

For a given rpm of the pump, a tube with larger inside diameter (ID) will give higher flow rate than one with a smaller inside diameter. Intuitively the flow rate is a function of the cross section area of the tube bore.

Flow rate

Flow rate is an important parameter for a pump. The flow rate in a peristaltic pump is determined by many factors, such as:

  1. Tube ID - higher flow rate with larger ID
  2. Pump head OD - higher flow rate with larger OD
  3. Pump head RPM - higher flow rate with higher RPM

Increasing the number of rollers doesn't increase the flow rate, instead it will decrease the flow rate somewhat by reducing the effective (i.e. fluid-pumping) circumference of the head. Increasing rollers does tend to decrease the amplitude of the fluid pulsing at the outlet by increasing the frequency of the pulsed flow.

The length of tube (measured from initial pinch point near the inlet to the final release point near the outlet) does not affect the flow rate. However, a longer tube implies more pinch points between inlet and outlet, increasing the pressure that the pump can generate.

Variations

Hose pumps

Higher pressure peristaltic hose pumps which can typically operate against up to 16 bar in continuous service, use shoes (rollers only used on low pressure types) and have casings filled with lubricant to prevent abrasion of the exterior of the pump tube and to aid in the dissipation of heat, and use reinforced tubes, often called "hoses". This class of pump is often called a "hose pump".

The biggest advantage with the hose pumps over the roller pumps is the high operating pressure of up to 16 bar. With rollers max pressure can arrive up to 12 Bar without any problem. If the high operating pressure is not required, a tubing pump is a better option than a hose pump if the pumped medium is not abrasive. With recent advances made in the tubing technology for pressure, life and chemical compatibility, as well as the higher flow rate ranges, the advantages that hose pumps had over roller pumps continues to erode.

Tube pumps

Lower pressure peristaltic pumps typically have dry casings and use rollers along with non-reinforced, extruded tubing. This class of pump is sometimes called a "tube pump" or "tubing pump". These pumps employ rollers to squeeze the tube. Except for the 360° eccentric pump design as described below, these pumps have a minimum of 2 rollers 180° apart, and may have as many as 8, or even 12 rollers. Increasing the number of rollers increase the pressure pulse frequency of the pumped fluid at the outlet, thereby decreasing the amplitude of pulsing. The downside to increasing number of rollers it that it proportionately increases number of squeezes, or occlusions, on the tubing for a given cumulative flow through that tube, thereby reducing the tubing life.

There are two kinds of roller design in peristaltic pumps:

1. Fixed occlusion - In this kind of pump, the rollers have a fixed locus as it turns, keeping the occlusion constant as it squeezes the tube. This is a simple, yet effective design. The only downside to this design is that the occlusion as a percent on the tube varies with the variation of the tube wall thickness. Typically the wall thickness of the extruded tubes vary enough that the % occlusion can vary with the wall thickness (see above). Therefore, a section of tube with greater wall thickness, but within the accepted tolerance, will have higher percent occlusion, which increases the wear on the tubing, thereby decreasing the tube life. Tube wall thickness tolerances today are generally kept tight enough that this issue is not of much practical concern. For those mechanically inclined, this may be the constant strain operation.

2. Spring-loaded rollers - As the name indicates, the rollers in this pump are mounted on a spring. This design is more elaborate than the fixed occlusion, but helps overcome the variations in the tube wall thickness over a broader range. Regardless of the variations, the roller imparts the same amount of stress on the tubing that is proportional to the spring constant, making this a constant stress operation. The spring is selected to overcome not only the hoop strength of the tubing, but also the pressure of the pumped fluid.

The operating pressure of these pumps is determined by the tubing and by the motor's ability to overcome the hoop strength of the tubing and the fluid pressure.

Advantages

  • No contamination. Because the only part of the pump in contact with the fluid being pumped is the interior of the tube, it is easy to sterilize and clean the inside surfaces of the pump.
  • Low maintenance needs. Their lack of valves, seals and glands makes them comparatively inexpensive to maintain.
  • They are able to handle slurries, viscous, shear-sensitive and aggressive fluids.
  • Pump design prevents backflow and syphoning without valves.[5]

Disadvantages

  • The flexible tubing will tend to degrade with time and require periodic replacement.
  • The flow is pulsed, particularly at low rotational speeds. Therefore, these pumps are less suitable where a smooth consistent flow is required. An alternative type of positive displacement pump should then be considered.

Tubing

It is important to select tubing with appropriate chemical resistance towards the liquid being pumped. Types of tubing commonly used in peristaltic pumps include:

Typical applications

Laboratory peristaltic pump
  • Agriculture
    • 'Sapsucker' pumps to extract maple tree sap
  • Food manufacturing and sales
    • Liquid food fountains
    • Beverage dispensing
  • Chemical handling
    • Printing, paint and pigments
    • Pharmaceutical production
    • Dosing systems for dishwasher and laundry chemicals
  • Engineering and manufacturing

References

  1. Dr. Michael E. DeBakey. "Methodist DeBakey Heart & Vascular Center". Retrieved 2010-06-27. 
  2. "Thermo Scientific Masterflex* Pumps Provide Superior Abrasive Handling, Significantly Lower Maintenance". 
  3. Handling Viscous Fluids
  4. How the ideal peristaltic pump should be constructed?
  5. Watson Marlow Pumps Group - How They Work , accessed 06-15-2010.

External links