Hounsfield scale

From Wikipedia, the free encyclopedia

The Hounsfield scale or CT numbers, named after Sir Godfrey Newbold Hounsfield, is a quantitative scale for describing radiodensity.

Definition

The Hounsfield unit (HU) scale is a linear transformation of the original linear attenuation coefficient measurement into one in which the radiodensity of distilled water at standard pressure and temperature (STP) is defined as zero Hounsfield units (HU), while the radiodensity of air at STP is defined as -1000 HU. In a voxel with average linear attenuation coefficient \mu _{X}, the corresponding HU value is therefore given by:

HU=1000\times {\frac  {\mu _{X}-\mu _{{water}}}{\mu _{{water}}}}

Where \mu _{{water}} is the linear attenuation coefficients of water.

Thus, a change of one Hounsfield unit (HU) represents a change of 0.1% of the attenuation coefficient of water since the attenuation coefficient of air is nearly zero.

It is the definition for CT scanners that are calibrated with reference to water.

Rationale

The above standards were chosen as they are universally available references and suited to the key application for which computed axial tomography was developed: imaging the internal anatomy of living creatures based on organized water structures and mostly living in air, e.g. humans.

The HU of common substances

The Hounsfield scale applies to medical grade CT scans but not to cone beam computed tomography (CBCT) scans.[1]

Substance HU
Air −1000
Lung −500
Fat −100 to −50
Water 0
CSF 15
Kidney 30
Blood +30 to +45
Muscle +10 to +40
Grey matter +37 to +45
White matter +20 to +30
Liver +40 to +60
J-O Blast +20 to +30
Soft Tissue, Contrast +100 to +300
Bone +700 (cancellous bone) to +3000 (dense bone)

A practical application of this is in evaluation of tumors, where, for example, an adrenal tumor with a radiodensity of less than 10 HU is rather fatty in composition and almost certainly a benign adrenal adenoma.[2]

History

CT machines were the first imaging devices for detailed visualization of the internal three-dimensional anatomy of living creatures, initially only as tomographic reconstructions of slice views or sections. Since the early 1990s, with advances in computer technology and scanners using spiral CT technology, internal three-dimensional anatomy is viewable by three-dimensional software reconstructions, from multiple perspectives, on computer monitors. By comparison, conventional X-ray images are two-dimensional projections of the true three-dimensional anatomy, i.e. radiodensity shadows. It was established by Sir Godfrey Newbold Hounsfield, one of the principal engineers and developers of computed axial tomography (CAT, or CT scans).

References

  • Feeman, Timothy G. (2010). The Mathematics of Medical Imaging: A Beginner's Guide. Springer Undergraduate Texts in Mathematics and Technology. Springer. ISBN 978-0387927114. 

Notes

  1. De Vos, W; et al. Cone-beam computerized tomography (CBCT) imaging of the oral and maxillofacial region: A systematic review of the literature. Int J Oral Maxillofac Surg 2009;38:609–625.
  2. medscape >Adrenal Adenoma Imaging. Author: Perry J Horwich. Chief Editor: Eugene C Lin. Updated: Apr 21, 2011

External links

This article is issued from Wikipedia. The text is available under the Creative Commons Attribution/Share Alike; additional terms may apply for the media files.