X-ray computed tomography

From Wikipedia, the free encyclopedia
X-ray computed tomography
Intervention
ICD-10-PCS B?2
ICD-9-CM 88.38
MeSH D014057
OPS-301 code: 3–20...3–26
MedlinePlus 003330

X-ray computed tomography (x-ray CT) is a technology that uses computer-processed x-rays to produce tomographic images (virtual 'slices') of specific areas of the scanned object, allowing the user to see what is inside it without cutting it open. Digital geometry processing is used to generate a three-dimensional image of the inside of an object from a large series of two-dimensional radiographic images taken around a single axis of rotation.[2] Medical imaging is the most common application of x-ray CT. Its cross-sectional images are used for diagnostic and therapeutic purposes in various medical disciplines.[3] The rest of this article discusses medical-imaging x-ray CT; industrial applications of x-ray CT are discussed at industrial computed tomography scanning.

Because x-ray CT is the most common form of CT in medicine and various other contexts, the term computed tomography alone (or CT) is often used to refer to x-ray CT, although other types exist (such as positron emission tomography [PET] and single-photon emission computed tomography [SPECT]). Older, less preferred terms that also refer to x-ray CT are computed axial tomography (CAT scan) and computer-assisted tomography. X-ray CT is a form of radiography, although the word "radiography" used alone usually refers, by wide convention, to non-tomographic radiography.

Sketch of a CT scanner.[1]

CT produces a volume of data that can be manipulated in order to demonstrate various bodily structures based on their ability to block the x-ray beam. Although, historically, the images generated were in the axial or transverse plane, perpendicular to the long axis of the body, modern scanners allow this volume of data to be reformatted in various planes or even as volumetric (3D) representations of structures. Although most common in medicine, CT is also used in other fields, such as nondestructive materials testing. Another example is archaeological uses such as imaging the contents of sarcophagi. Individuals responsible for performing CT exams are called radiologic technologists or radiographers[4] and are required to be licensed in most states of the USA.[5]

Usage of CT has increased dramatically over the last two decades in many countries.[6] An estimated 72 million scans were performed in the United States in 2007.[7] One study estimated that as many as 0.4% of current cancers in the United States are due to CTs performed in the past and that this may increase to as high as 1.5 to 2% with 2007 rates of CT usage;[8] however, this estimate is disputed.[9] Kidney problems following intravenous contrast agents may also be a concern in some types of studies.

Diagnostic use

Picture of a CT scout (aka scanogram or topogram) as used for planning every scan slice.

Since its introduction in the 1970s, CT has become an important tool in medical imaging to supplement x-rays and medical ultrasonography. It has more recently been used for preventive medicine or screening for disease, for example CT colonography for patients with a high risk of colon cancer, or full-motion heart scans for patients with high risk of heart disease. A number of institutions offer full-body scans for the general population although this practice goes against the advice and official position of many professional organizations in the field.[10]

Head

Computed tomography of human brain, from base of the skull to top. Taken with intravenous contrast medium.
Bone reconstructed in 3D

CT scanning of the head is typically used to detect infarction, tumors, calcifications, haemorrhage and bone trauma. Of the above, hypodense (dark) structures can indicate edema and infarction, hyperdense (bright) structures indicate calcifications and haemorrhage and bone trauma can be seen as disjunction in bone windows. Tumors can be detected by the swelling and anatomical distortion they cause, or by surrounding edema. Ambulances equipped with small bore multi-sliced CT scanners respond to cases involving stroke or head trauma.

Magnetic resonance imaging (MRI) of the head provides superior information as compared to CT scans when seeking information about headache to confirm a diagnosis of neoplasm, vascular disease, posterior cranial fossa lesions, cervicomedullary lesions, or intracranial pressure disorders.[11] It also does not carry the risks of exposing the patient to ionizing radiation.[11] CT scans may be used to diagnose headache when neuroimaging is indicated and MRI is not available, or in emergency settings when hemorrhage, stroke, or traumatic brain injury are suspected.[11]

Lungs

CT can be used for detecting both acute and chronic changes in the lung parenchyma, that is, the internals of the lungs. It is particularly relevant here because normal two-dimensional X-rays do not show such defects. A variety of techniques are used, depending on the suspected abnormality. For evaluation of chronic interstitial processes (emphysema, fibrosis, and so forth), thin sections with high spatial frequency reconstructions are used; often scans are performed both in inspiration and expiration. This special technique is called high resolution CT. Therefore, it produces a sampling of the lung and not continuous images.

An incidentally found nodule in the absence of symptoms (sometimes referred to as an incidentaloma) may raise concerns that it might represent a tumor, either benign or malignant.[12] Perhaps persuaded by fear, patients and doctors sometimes agree to an intensive schedule of CT scans, sometimes up to every three months and beyond the recommended guidelines, in an attempt to do surveillance on the nodules.[13] However, established guidelines advise that patients without a prior history of cancer and whose solid nodules have not grown over a two-year period are unlikely to have any malignant cancer.[13] For this reason, and because no research provides supporting evidence that intensive surveillance gives better outcomes, and because of risks associated with having CT scans, patients should not receive CT screening in excess of those recommended by established guidelines.[13]

Pulmonary angiogram

Example of a CTPA, demonstrating a saddle embolus (dark horizontal line) occluding the pulmonary arteries (bright white triangle)

CT pulmonary angiogram (CTPA) is a medical diagnostic test used to diagnose pulmonary embolism (PE). It employs computed tomography and an iodine based contrast agent to obtain an image of the pulmonary arteries.

Cardiac

With the advent of subsecond rotation combined with multi-slice CT (up to 320-slices), high resolution and high speed can be obtained at the same time, allowing excellent imaging of the coronary arteries (cardiac CT angiography).

Abdominal and pelvic

CT Scan of 11 cm Wilms' tumor of right kidney in 13 month old patient.

CT is a sensitive method for diagnosis of abdominal diseases. It is used frequently to determine stage of cancer and to follow progress. It is also a useful test to investigate acute abdominal pain.

Extremities

CT is often used to image complex fractures, especially ones around joints, because of its ability to reconstruct the area of interest in multiple planes. Fractures, ligamentous injuries and dislocations can easily be recognised with a 0.2 mm resolution.[14][15]

Advantages

There are several advantages that CT has over traditional 2D medical radiography. First, CT completely eliminates the superimposition of images of structures outside the area of interest. Second, because of the inherent high-contrast resolution of CT, differences between tissues that differ in physical density by less than 1% can be distinguished. Finally, data from a single CT imaging procedure consisting of either multiple contiguous or one helical scan can be viewed as images in the axial, coronal, or sagittal planes, depending on the diagnostic task. This is referred to as multiplanar reformatted imaging.

CT is regarded as a moderate- to high-radiation diagnostic technique. The improved resolution of CT has permitted the development of new investigations, which may have advantages; compared to conventional radiography, for example, CT angiography avoids the invasive insertion of a catheter. CT Colonography (also known as Virtual Colonoscopy or VC for short) may be as useful as a barium enema for detection of tumors, but may use a lower radiation dose. CT VC is increasingly being used in the UK as a diagnostic test for bowel cancer and can negate the need for a colonoscopy.

The radiation dose for a particular study depends on multiple factors: volume scanned, patient build, number and type of scan sequences, and desired resolution and image quality. In addition, two helical CT scanning parameters that can be adjusted easily and that have a profound effect on radiation dose are tube current and pitch. Computed tomography (CT) scan has been shown to be more accurate than radiographs in evaluating anterior interbody fusion but may still over-read the extent of fusion.[16]

Adverse effects

Cancer

The radiation used in CT scans can damage body cells, including DNA molecules, which can lead to cancer.[8] A single CT scan will expose a patient to enough radiation to cause cancer, while children who received multiple CT scans were found to be three times more likely to develop leukemia and brain cancer.[17] The Institute of Medicine found that such imaging radiation, along with hormone therapy, were, until recent years, the "leading" causes of breast cancer."[17] The most common cancers caused by radiation exposure are thought to be lung cancer, breast cancer, thyroid cancer, stomach cancer and leukemia.[18]

According to the National Council on Radiation Protection and Measurements, between the 1980s and 2006, the use of CT scans has increased sixfold (600%). The radiation doses received from CT scans are 100 to 1,000 times higher than conventional X-rays.[17] A study by a New York hospital found that nearly a third of its patients who underwent multiple scans received the equivalent of 5,000 chest X-rays.[17]

Some experts note that CT scans are known to be "overused," and "there is distressingly little evidence of better health outcomes associated with the current high rate of scans."[17]

Early estimates of harm from CT are partly based on similar radiation exposures experienced by those present during the atomic bomb explosions in Japan during the second world war and those of nuclear industry workers.[8] A more recent study by the National Cancer Institute in 2009, based on scans made in 2007, estimated that 29,000 excess cancer cases and 14,500 excess deaths would be caused over the lifetime of the patients. Some experts project that in the future, between three and five percent of all cancers would result from medical imaging.[17]

A person's age plays a significant role in the subsequent risk of cancer.[19] Estimated lifetime cancer mortality risks from an abdominal CT of a 1-year-old is 0.1% or 1:1000 scans.[19] The risk for someone who is 40 years old is half that of someone who is 20 years old with substantially less risk in the elderly.[19] The International Commission on Radiological Protection estimates that the risk to a fetus being exposed to 10 mGy (a unit of radiation exposure, see Gray (unit)) increases the rate of cancer before 20 years of age from 0.03% to 0.04% (for reference a CT pulmonary angiogram exposes a fetus to 4 mGy).[20] A 2012 review did not find an association between medical radiation and cancer risk in children noting however the existence of limitations in the evidences over which the review is based.[21]

CT scans can be performed with different settings for lower exposure in children with most manufacturers of CT scans as of 2007 having this function built in.[18] Furthermore, certain conditions can require children to be exposed to multiple CT scans.[8] Studies support informing parents of the risks of pediatric CT scanning.[22]

Contrast

In the United States half of CT scans involve intravenously injected radiocontrast agents.[23] The most common reactions from these agents are mild, including nausea, vomiting and an itching rash; however, more severe reactions may occur.[24] Overall reactions occur in 1 to 3% with nonionic contrast and 4 to 12% of people with ionic contrast.[25] Skin rashes may appear within a week to 3% of people.[24]

The old radiocontrast agents caused anaphylaxis in 1% of cases while the newer, lower-osmolar agents cause reactions in 0.01–0.04% of cases.[24][26] Death occurs in about two to 30 people per 1,000,000 administrations with newer agents being safer.[25][27] When deaths do occur it is more typically in those who are female, elderly or in poor health and is secondary to either anaphylaxis or acute renal failure.[23]

The contrast agent may induce contrast-induced nephropathy.[28] This occurs in 2 to 7% of people who receive these agents, with greater risk in those who have preexisting renal insufficiency,[28] preexisting diabetes, or reduced intravascular volume. People with mild kidney impairment are usually advised to ensure full hydration for several hours before and after the injection. For moderate kidney failure, the use of iodinated contrast should be avoided; this may mean using an alternative technique instead of CT. Those with severe renal failure requiring dialysis require less strict precautions, as their kidneys have so little function remaining that any further damage would not be noticeable and the dialysis will remove the contrast agent; it is normally recommended, however, to arrange dialysis as soon as possible following contrast administration to minimize any adverse effects of the contrast.

In addition to the use of intravenous contrast, orally administered contrast agents are frequently used when examining the abdomen. These are frequently the same as the intravenous contrast agents, merely diluted to approximately 10% of the concentration. However, oral alternatives to iodinated contrast exist, such as very dilute (0.5–1% w/v) barium sulfate suspensions. Dilute barium sulfate has the advantage that it does not cause allergic-type reactions or kidney failure, but cannot be used in patients with suspected bowel perforation or suspected bowel injury, as leakage of barium sulfate from damaged bowel can cause fatal peritonitis.

Scan dose

Examination Typical effective
dose (mSv)
to the whole body
Typical absorbed
dose
(mGy)
to the organ in question
Annual background radiation 2.4[29] 2.4[29]
Chest X-ray 0.02[30] 0.01–0.15[31]
Head CT 1–2[19] 56[32]
Screening mammography 0.4[20] 3[8][31]
Abdomen CT 8[30] 14[32]
Chest CT 5–7[19] 13[32]
CT colonography 6–11[19]
Chest, abdomen and pelvis CT 9.9[32] 12[32]
Cardiac CT angiogram 9–12[19] 40–100[31]
Barium enema 15[8] 15[31]
Neonatal abdominal CT 20[8] 20[31]

The table reports average radiation exposures, however, there can be a wide variation in radiation doses between similar scan types, where the highest dose could be as much as 22 times higher than the lowest dose.[19] A typical plain film x-ray involves radiation dose of 0.01 to 0.15 mGy, while a typical CT can involve 10–20 mGy for specific organs, and can go up to 80 mGy for certain specialized CT scans.[31]

For purposes of comparison, the world average dose rate from naturally occurring sources of background radiation is 2.4 mSv per year, equal for practical purposes in this application to 2.4 mGy per year.[29] While there is some variation, most people (99%) received less than 7 mSv per year as background radiation.[33] Medical imaging as of 2007 accounted for half of the radiation exposure of those in the United States with CT scans making up two thirds of this amount.[19] In the United Kingdom it accounts for 15% of radiation exposure.[20] The average radiation dose from medical sources is ~0.6 mSv per person globally as of 2007.[19] Those in the nuclear industry in the United States are limited to doses of 50 mSv a year and 100 mSv every 5 years.[19]

Radiation dose units

The radiation dose reported in the gray or mGy unit is proportional to the amount of energy that the irradiated body part is expected to absorb, and the physical effect (such as DNA double strand breaks) on the cells' chemical bonds by x-ray radiation is proportional to that energy.[34]

The sievert unit is used in the report of the effective dose. The sievert unit in the context of CT scans, does not correspond to the actual radiation dose that the scanned body part absorbs, but rather to another radiation dose of another scenario, in which the whole body absorbs the other radiation dose, and where the other radiation dose is of a magnitude that is estimated to have the same probability to induce cancer as the CT scan.[35] Thus, as is shown in the table above, the actual radiation that is absorbed by a scanned body part is often much larger than the effective dose suggests. A specific measure, termed the computed tomography dose index (CTDI), is commonly used as an estimate of the radiation absorbed dose for tissue within the scan region, and is automatically computed by medical CT scanners.

The equivalent dose is the effective dose of a case, in which the whole body would actually absorb the same radiation dose, and the sievert unit is used in its report. In the case of non-uniform radiation, or radiation given to only part of the body, which is common for CT examinations, using the local equivalent dose alone would overstate the biological risks to the entire organism.

Excess doses

In October, 2009, the US Food and Drug Administration (FDA) initiated an investigation of brain perfusion CT (PCT) scans, based on overdoses of radiation caused by incorrect settings at one particular facility for this particular type of CT scan. Over 256 patients over an 18 month period were exposed, over 40% lost patches of hair, and prompted the editorial to call for increased CT quality assurance programs, while also noting that "while unnecessary radiation exposure should be avoided, a medically needed CT scan obtained with appropriate acquisition parameter has benefits that outweigh the radiation risks."[19][36] Similar problems have been reported at other centers.[19] These incidents are believed to be due to human error.[19]

Campaigns

In response to increased concern by the public and the ongoing progress of best practices, The Alliance for Radiation Safety in Pediatric Imaging was formed within the Society for Pediatric Radiology. In concert with The American Society of Radiologic Technologists, The American College of Radiology and The American Association of Physicists in Medicine, the Society for Pediatric Radiology developed and launched the Image Gently Campaign which is designed to maintain high quality imaging studies while using the lowest doses and best radiation safety practices available on pediatric patients.[37] This initiative has been endorsed and applied by a growing list of various professional medical organizations around the world and has received support and assistance from companies that manufacture equipment used in Radiology.

Following upon the success of the Image Gently campaign, the American College of Radiology, the Radiological Society of North America, the American Association of Physicists in Medicine and the American Society of Radiologic Technologists have launched a similar campaign to address this issue in the adult population called Image Wisely.[38]

The World Health Organization and International Atomic Energy Agency (IAEA) of the United Nations have also been working in this area and have ongoing projects designed to broaden best practices and lower patient radiation dose.[39][40]

Prevalence

Usage of CT has increased dramatically over the last two decades.[6] An estimated 72 million scans were performed in the United States in 2007.[7] Of these, six to eleven percent are done in children,[20] an increase of seven to eightfold from 1980.[19] Similar increases have been seen in Europe and Asia.[19] In Calgary, Canada 12.1% of people who present to the emergency with an urgent complaint received a CT scan, most commonly either of the head or of the abdomen. The percentage who received CT, however, varied markedly by the emergency physician who saw them from 1.8% to 25%.[41] In the emergency department in the United States, CT or MRI imaging is done in 15% of people who present with injuries as of 2007 (up from 6% in 1998).[42]

The increased use of CT scans has been the greatest in two fields: screening of adults (screening CT of the lung in smokers, virtual colonoscopy, CT cardiac screening, and whole-body CT in asymptomatic patients) and CT imaging of children. Shortening of the scanning time to around 1 second, eliminating the strict need for the subject to remain still or be sedated, is one of the main reasons for the large increase in the pediatric population (especially for the diagnosis of appendicitis).[8] As of 2007 in the United States a proportion of CT scans are performed unnecessarily.[18] Some estimates place this number at 30%.[20] There are a number of reasons for this including: legal concerns, financial incentives, and desire by the public.[18] There is also otherwise healthy people receiving full body CT scans as screening for which there is no evidence to support.[18]

Process

3D reconstruction of the brain and eyes from CT scanned DICOM images. In this image, areas with the density of bone or air were made transparent, and the slices stacked up in an approximate free-space alignment. The outer ring of material around the brain are the soft tissues of skin and muscle on the outside of the skull. A black box encloses the slices to provide the black background. Since these are simply 2D images stacked up, when viewed on edge the slices disappear since they have effectively zero thickness. Each DICOM scan represents about 5mm of material averaged into a thin slice.

X-ray slice data is generated using an X-ray source that rotates around the object; X-ray sensors are positioned on the opposite side of the circle from the X-ray source. The earliest sensors were scintillation detectors, with photomultiplier tubes excited by (typically) cesium iodide crystals. Cesium iodide was replaced during the 1980s by ion chambers containing high-pressure Xenon gas. These systems were in turn replaced by scintillation systems based on photodiodes instead of photomultipliers and modern scintillation materials (for example rare earth garnet or rare earth oxide ceramics) with more desirable characteristics.

Initial machines would rotate the X-ray source and detectors around a stationary object. Following a complete rotation, the object would be moved along its axis, and the next rotation started. Newer machines permitted continuous rotation with the object to be imaged slowly and smoothly slid through the X-ray ring. These are called helical or spiral CT machines. A subsequent development of helical CT was multi-slice (or multi-detector) CT; instead of a single row of detectors, multiple rows of detectors are used effectively capturing multiple cross-sections simultaneously. Systems with a very large number of detector rows, such that the z-axis coverage is comparable to the xy-axis coverage are often termed cone beam CT, due to the shape of the X-ray beam (strictly, the beam is pyramidal in shape, rather than conical).

In conventional CT machines, an X-ray tube and detector are physically rotated behind a circular shroud (see the image above right). An alternative, short lived design, known as electron beam tomography (EBT), used electromagnetic deflection of an electron beam within a very large conical X-ray tube and a stationary array of detectors to achieve very high temporal resolution, for imaging of rapidly moving structures, for example the coronary arteries. Cone-beam CT functionality is also an increasingly common function found in medical fluoroscopy equipment; by rotating the fluoroscope around the patient, a geometry similar to CT can be obtained, and by treating the 2D X-ray detector in a manner similar to a CT detector with a massive number of rows, it is possible to reconstruct a 3D volume from a single rotation using suitable software.

CT scanner with cover removed to show internal components. Legend:
T: X-ray tube
D: X-ray detectors
X: X-ray beam
R: Gantry rotation

CT is used in medicine as a diagnostic tool and as a guide for interventional procedures. Sometimes contrast materials such as intravenous iodinated contrast are used. This is useful to highlight structures such as blood vessels that otherwise would be difficult to delineate from their surroundings. Using contrast material can also help to obtain functional information about tissues.

A visual representation of the raw data obtained is called a sinogram, yet it is not sufficient for interpretation. Once the scan data has been acquired, the data must be processed using a form of tomographic reconstruction, which produces a series of cross-sectional images. In terms of mathematics, the raw data acquired by the scanner consists of multiple "projections" of the object being scanned. These projections are effectively the Radon transformation of the structure of the object. Reconstruction, essentially involves solving the inverse Radon transformation.

The technique of filtered back projection is one of the most established algorithmic techniques for this problem. It is conceptually simple, tunable and deterministic. It is also computationally undemanding, with modern scanners requiring only a few milliseconds per image. However, this is not the only technique available: the original EMI scanner solved the tomographic reconstruction problem by linear algebra, but this approach was limited by its high computational complexity, especially given the computer technology available at the time. More recently, manufacturers have developed iterative physical model-based maximum likelihood expectation maximization techniques. These techniques are advantageous because they use an internal model of the scanner's physical properties and of the physical laws of X-ray interactions. Earlier methods, such as filtered back projection, assume a perfect scanner and highly simplified physics, which leads to a number of artifacts, high noise and impaired image resolution. Iterative techniques provide images with improved resolution, reduced noise and fewer artifacts, as well as the ability to greatly reduce the radiation dose in certain circumstances. The disadvantage is a very high computational requirement, but advances in computer technology and high-performance computing techniques, such as use of highly parallel GPU algorithms, now allow practical use.

Pixels in an image obtained by CT scanning are displayed in terms of relative radiodensity. The pixel itself is displayed according to the mean attenuation of the tissue(s) that it corresponds to on a scale from +3071 (most attenuating) to −1024 (least attenuating) on the Hounsfield scale. Pixel is a two dimensional unit based on the matrix size and the field of view. When the CT slice thickness is also factored in, the unit is known as a Voxel, which is a three-dimensional unit. The phenomenon that one part of the detector cannot differentiate between different tissues is called the "Partial Volume Effect". That means that a big amount of cartilage and a thin layer of compact bone can cause the same attenuation in a voxel as hyperdense cartilage alone. Water has an attenuation of 0 Hounsfield units (HU), while air is −1000 HU, cancellous bone is typically +400 HU, cranial bone can reach 2000 HU or more (os temporale) and can cause artifacts. The attenuation of metallic implants depends on atomic number of the element used: Titanium usually has an amount of +1000 HU, iron steel can completely extinguish the X-ray and is, therefore, responsible for well-known line-artifacts in computed tomograms. Artifacts are caused by abrupt transitions between low- and high-density materials, which results in data values that exceed the dynamic range of the processing electronics.

Contrast mediums used for X-ray CT, as well as for plain film X-ray, are called radiocontrasts. Radiocontrasts for X-ray CT are, in general, iodine-based.[43] Often, images are taken both with and without radiocontrast. CT images are called precontrast or native-phase images before any radiocontrast has been administrated, and postcontrast after radiocontrast administration.[44]

Two-dimensional CT images are conventionally rendered so that the view is as though looking up at it from the patient's feet.[45] Hence, the left side of the image is to the patient's right and vice versa, while anterior in the image also is the patient's anterior and vice versa. This left-right interchange corresponds to the view that physicians generally have in reality when positioned in front of patients.

CT data sets have a very high dynamic range which must be reduced for display or printing. This is typically done via a process of "windowing", which maps a range (the "window") of pixel values to a greyscale ramp. For example, CT images of the brain are commonly viewed with a window extending from 0 HU to 80 HU. Pixel values of 0 and lower, are displayed as black; values of 80 and higher are displayed as white; values within the window are displayed as a grey intensity proportional to position within the window. The window used for display must be matched to the X-ray density of the object of interest, in order to optimize the visible detail.

Three-dimensional reconstruction

Because contemporary CT scanners offer isotropic or near isotropic, resolution, display of images does not need to be restricted to the conventional axial images. Instead, it is possible for a software program to build a volume by "stacking" the individual slices one on top of the other. The program may then display the volume in an alternative manner.[46]

Multiplanar reconstruction

Typical screen layout for diagnostic software, showing one 3D and three MPR views

Multiplanar reconstruction (MPR) is the simplest method of reconstruction. A volume is built by stacking the axial slices. The software then cuts slices through the volume in a different plane (usually orthogonal). As an option, a special projection method, such as maximum-intensity projection (MIP) or minimum-intensity projection (mIP/MinIP), can be used to build the reconstructed slices.

MPR is frequently used for examining the spine. Axial images through the spine will only show one vertebral body at a time and cannot reliably show the intervertebral discs. By reformatting the volume, it becomes much easier to visualise the position of one vertebral body in relation to the others.

Modern software allows reconstruction in non-orthogonal (oblique) planes so that the optimal plane can be chosen to display an anatomical structure. This may be particularly useful for visualising the structure of the bronchi as these do not lie orthogonal to the direction of the scan.

For vascular imaging, curved-plane reconstruction can be performed. This allows bends in a vessel to be "straightened" so that the entire length can be visualised on one image, or a short series of images. Once a vessel has been "straightened" in this way, quantitative measurements of length and cross sectional area can be made, so that surgery or interventional treatment can be planned.

MIP reconstructions enhance areas of high radiodensity, and so are useful for angiographic studies. MIP reconstructions tend to enhance air spaces so are useful for assessing lung structure.

3D rendering techniques

Surface rendering of the head of the frog Atelopus franciscus, with ear parts highlighted.

Surface rendering

A threshold value of radiodensity is set by the operator (e.g., a level that corresponds to bone). From this, a three-dimensional model can be constructed using edge detection image processing algorithms and displayed on screen. Multiple models can be constructed from various thresholds, allowing different colors to represent each anatomical component such as bone, muscle, and cartilage. However, the interior structure of each element is not visible in this mode of operation.

Volume rendering

Surface rendering is limited in that it will display only surfaces that meet a threshold density, and will display only the surface that is closest to the imaginary viewer. In volume rendering, transparency, colors and shading are used to allow a better representation of the volume to be shown in a single image. For example, the bones of the pelvis could be displayed as semi-transparent, so that, even at an oblique angle, one part of the image does not conceal another.

Image segmentation

Where different structures have similar radiodensity, it can become impossible to separate them simply by adjusting volume rendering parameters. The solution is called segmentation, a manual or automatic procedure that can remove the unwanted structures from the image.

Image quality

Artifacts

Although images produced by CT are generally faithful representations of the scanned volume, the technique is susceptible to a number of artifacts, such as the following:[2][47]Chapters 3 and 5

Streak artifact
Streaks are often seen around materials that block most X-rays, such as metal or bone. Numerous factors contribute to these streaks: undersampling, photon starvation, motion, beam hardening, and Compton scatter. This type of artifact commonly occurs in the posterior fossa of the brain, or if there are metal implants. The streaks can be reduced using newer reconstruction techniques[48] or approaches such as metal artifact reduction (MAR).[49]
Partial volume effect
This appears as "blurring" of edges. It is due to the scanner being unable to differentiate between a small amount of high-density material (e.g., bone) and a larger amount of lower density (e.g., cartilage). The reconstruction assumes that the X-ray attenuation within each voxel is homogenous; this may not be the case at sharp edges. This is most commonly seen in the z-direction, due to the conventional use of highly anisotropic voxels, which have a much lower out-of-plane resolution, than in-plane resolution. This can be partially overcome by scanning using thinner slices, or an isotropic acquisition on a modern scanner.
Ring artifact
Probably the most common mechanical artifact, the image of one or many "rings" appears within an image. This is usually due to a detector fault, or miscalibration of an individual detector element.
Noise
This appears as grain on the image and is caused by a low signal to noise ratio. This occurs more commonly when a thin slice thickness is used. It can also occur when the power supplied to the X-ray tube is insufficient to penetrate the anatomy.
Motion artifact
This is seen as blurring and/or streaking, which is caused by movement of the object being imaged. Motion blurring might be reduced using a new technique called IFT (incompressible flow tomography).[50]
Windmill
Streaking appearances can occur when the detectors intersect the reconstruction plane. This can be reduced with filters or a reduction in pitch.
Beam hardening
This can give a "cupped appearance". It occurs when there is more attenuation along a path passing through the center of an object, than a path that grazes the edge. This is easily corrected by filtration and software.[51][52]

Dose vs. image quality

An important issue within radiology today is how to reduce the radiation dose during CT examinations without compromising the image quality. In general, higher radiation doses result in higher-resolution images, while lower doses lead to increased image noise and unsharp images. However, increased dosage raises increase the adverse side effects, including the risk of radiation induced cancer – a four-phase abdominal CT gives the same radiation dose as 300 chest x-rays (See the Scan dose section). Several methods that can reduce the exposure to ionizing radiation during a CT scan exist.

  1. New software technology can significantly reduce the required radiation dose.
  2. Individualize the examination and adjust the radiation dose to the body type and body organ examined. Different body types and organs require different amounts of radiation.
  3. Prior to every CT examination, evaluate the appropriateness of the exam whether it is motivated or if another type of examination is more suitable. Higher resolution is not always suitable for any given scenario, such as detection of small pulmonary masses.[53]

Industrial use

Industrial CT Scanning (industrial computed tomography) is a process which utilizes x-ray equipment to produce 3D representations of components both externally and internally. Industrial CT scanning has been utilized in many areas of industry for internal inspection of components. Some of the key uses for CT scanning have been flaw detection, failure analysis, metrology, assembly analysis, and reverse engineering applications. CT scanning is also employed in the imaging and conservation of museum artifacts.[54]

CT scanning has also found an application in transport security (predominantly airport security where it is currently used in a materials analysis context for explosives detection CTX (explosive-detection device) and is also under consideration for automated baggage/parcel security scanning using computer vision based object recognition algorithms that target the detection of specific threat items based on 3D appearance (e.g. guns, knives, liquid containers).[55][56][57]

History

The prototype CT scanner
A historic EMI-Scanner

Origins of tomography

In the early 1900s, the Italian radiologist Alessandro Vallebona proposed a method to represent a single slice of the body on radiographic film. This method was known as tomography. The idea is based on simple principles of projective geometry: moving synchronously and in opposite directions the X-ray tube and the film, which are connected together by a rod whose pivot point is the focus; the image created by the points on the focal plane appears sharper, while the images of the other points annihilate as noise.[58] This is only marginally effective, as blurring occurs in only the "x" plane. Now known as conventional tomography, this method of acquiring tomogaphic images using only mechanical techniques advanced through the mid-twentieth century, steadily producing sharper images, and with a greater ability to vary the thickness of the cross-section being examined. This was achieved through the introduction of more complex, pluridirectional devices that can move in more than one plane and perform more effective blurring. However, despite the increasing sophistication of conventional tomography, it remained ineffective at producing images of soft tissues.[58] With the increasing power and availability of computers in the 1960s, research began into practical computational techniques for creating tomographic images.

Mathematical theory

The mathematical theory behind the tomographic reconstruction dates back to 1917 where the invention of Radon Transform[59][60] by an Austrian mathematician Johann Radon. He showed mathematically that a function could be reconstructed from an infinite set of its projections.[61] In 1937, a Polish mathematician, named Stefan Kaczmarz, developed a method to find an approximate solution to a large system of linear algebraic equations.[62][63] This led the foundation to another powerful reconstruction method called "Algebraic Reconstruction Technique (ART)" which was later adapted by Sir Godfrey Hounsfield as the image reconstruction mechanism in his famous invention, the first commercial CT scanner.

In 1956, Ronald N. Bracewell used a method similar to the Radon Transform to reconstruct a map of solar radiation from a set of solar radiation measurements.[64] In 1959, William Oldendorf, a UCLA neurologist and senior medical investigator at the West Los Angeles Veterans Administration hospital, conceived an idea for "scanning a head through a transmitted beam of X-rays, and being able to reconstruct the radiodensity patterns of a plane through the head" after watching an automated apparatus built to reject frostbitten fruit by detecting dehydrated portions. In 1961, he built a prototype in which an X-ray source and a mechanically coupled detector rotated around the object to be imaged. By reconstructing the image, this instrument could get an X-ray picture of a nail surrounded by a circle of other nails, which made it impossible to X-ray from any single angle.[65] In his landmark paper published in 1961, he described the basic concept which was later used by Allan McLeod Cormack to develop the mathematics behind computerized tomography.

In October 1963, Oldendorf received a U.S. patent for a "radiant energy apparatus for investigating selected areas of interior objects obscured by dense material." Oldendorf shared the 1975 Lasker award with Hounsfield for that discovery.[66] The field of the mathematical methods of computerized tomography has seen a very active development since then, as is evident from overview literature[2][67][68] by Frank Natterer and Gabor T. Herman, two of the pioneers in the this field.[69]

Tomography has been one of the pillars of radiologic diagnostics until the late 1970s, when the availability of minicomputers and of the transverse axial scanning method led CT to gradually supplant conventional tomography as the preferred modality of obtaining tomographic images. Transverse axial scanning was due in large part to the work of Godfrey Hounsfield and South African-born Allan McLeod Cormack. In terms of mathematics, the method is based upon the use of the Radon Transform. But as Cormack remembered later,[70] he had to find the solution himself since it was only in 1972 that he learned of the work of Radon, by chance.

Commercial scanners

The first commercially viable CT scanner was invented by Sir Godfrey Hounsfield in Hayes, United Kingdom, at EMI Central Research Laboratories using X-rays. Hounsfield conceived his idea in 1967.[71] The first EMI-Scanner was installed in Atkinson Morley Hospital in Wimbledon, England, and the first patient brain-scan was done on 1 October 1971.[72] It was publicly announced in 1972.

The original 1971 prototype took 160 parallel readings through 180 angles, each 1° apart, with each scan taking a little over 5 minutes. The images from these scans took 2.5 hours to be processed by algebraic reconstruction techniques on a large computer. The scanner had a single photomultiplier detector, and operated on the Translate/Rotate principle.[72]

It is often claimed that revenues from the sales of The Beatles records in the 1960s helped funding the development of the first CT scanner at EMI[73] although this has recently been disputed.[74] The first production X-ray CT machine (in fact called the "EMI-Scanner") was limited to making tomographic sections of the brain, but acquired the image data in about 4 minutes (scanning two adjacent slices), and the computation time (using a Data General Nova minicomputer) was about 7 minutes per picture. This scanner required the use of a water-filled Perspex tank with a pre-shaped rubber "head-cap" at the front, which enclosed the patient's head. The water-tank was used to reduce the dynamic range of the radiation reaching the detectors (between scanning outside the head compared with scanning through the bone of the skull). The images were relatively low resolution, being composed of a matrix of only 80 × 80 pixels.

In the U.S., the first installation was at the Mayo Clinic. As a tribute to the impact of this system on medical imaging the Mayo Clinic has an EMI scanner on display in the Radiology Department. Allan McLeod Cormack of Tufts University in Massachusetts independently invented a similar process, and both Hounsfield and Cormack shared the 1979 Nobel Prize in Medicine.[75][76]

The first CT system that could make images of any part of the body and did not require the "water tank" was the ACTA (Automatic Computerized Transverse Axial) scanner designed by Robert S. Ledley, DDS, at Georgetown University. This machine had 30 photomultiplier tubes as detectors and completed a scan in only nine translate/rotate cycles, much faster than the EMI-Scanner. It used a DEC PDP11/34 minicomputer both to operate the servo-mechanisms and to acquire and process the images. The Pfizer drug company acquired the prototype from the university, along with rights to manufacture it. Pfizer then began making copies of the prototype, calling it the "200FS" (FS meaning Fast Scan), which were selling as fast as they could make them. This unit produced images in a 256×256 matrix, with much better definition than the EMI-Scanner's 80×80.

Since the first CT scanner, CT technology has vastly improved. Improvements in speed, slice count, and image quality have been the major focus primarily for cardiac imaging. Scanners now produce images much faster and with higher resolution enabling doctors to diagnose patients more accurately and perform medical procedures with greater precision. In the late 1990s CT scanners broke into two major groups, "Fixed CT" and "Portable CT". "Fixed CT Scanners" are large, require a dedicated power supply, electrical closet, HVAC system, a separate workstation room, and a large lead lined room. "Fixed CT Scanners" can also be mounted inside large tractor trailers and driven from site to site and are known as "Mobile CT Scanners". "Portable CT Scanners" are light weight, small, and mounted on wheels. These scanners often have built-in lead shielding and run off of batteries or standard wall power.

In 2013, General Electric was awaiting approval for a new generation of scanners that can take an image in less than 1 second, fast enough to produce clear images of beating hearts and coronary arteries.[77]

Etymology

The word "tomography" is derived from the Greek tomos (slice) and graphein (to write). Computed tomography was originally known as the "EMI scan" as it was developed in the early 1970s at a research branch of EMI, a company best known today for its music and recording business. It was later known as computed axial tomography (CAT or CT scan) and body section röntgenography.

Although the term "computed tomography" could be used to describe positron emission tomography or single photon emission computed tomography (SPECT), in practice it usually refers to the computation of tomography from X-ray images, especially in older medical literature and smaller medical facilities.

In MeSH, "computed axial tomography" was used from 1977 to 1979, but the current indexing explicitly includes "X-ray" in the title.[78]

Types of machines

Spinning tube, commonly called spiral CT, or helical CT in which an entire X-ray tube is spun around the central axis of the area being scanned. These are the dominant type of scanners on the market because they have been manufactured longer and offer lower cost of production and purchase. The main limitation of this type is the bulk and inertia of the equipment (X-ray tube assembly and detector array on the opposite side of the circle) which limits the speed at which the equipment can spin. Some designs use two X-ray sources and detector arrays offset by an angle, as technique to improve temporal resolution.

Electron beam tomography (EBT) is a specific form of CT in which a large enough X-ray tube is constructed so that only the path of the electrons, travelling between the cathode and anode of the X-ray tube, are spun using deflection coils. This type had a major advantage since sweep speeds can be much faster, allowing for less blurry imaging of moving structures, such as the heart and arteries. Fewer scanners of this design have been produced when compared with spinning tube types, mainly due to the higher cost associated with building a much larger X-ray tube and detector array and limited anatomical coverage. Only one manufacturer (Imatron, later acquired by General electric) ever produced scanners of this design. Production ceased in early 2006.[79]

Previous studies

Pneumoencephalography of the brain was quickly replaced by CT. A form of tomography can be performed by moving the X-ray source and detector during an exposure. Anatomy at the target level remains sharp, while structures at different levels are blurred. By varying the extent and path of motion, a variety of effects can be obtained, with variable depth of field and different degrees of blurring of "out of plane" structures.[80]:25 Although largely obsolete, conventional tomography is still used in specific situations such as dental imaging (orthopantomography) or in intravenous urography.

See also

References

  1. US Patent 4115698 
  2. 2.0 2.1 2.2 Herman, G. T., Fundamentals of computerized tomography: Image reconstruction from projection, 2nd edition, Springer, 2009
  3. "computed tomography – Definition from the Merriam-Webster Online Dictionary". Retrieved 2009-08-18. 
  4. "Patient Page | ARRT - The American Registry of Radiologic Technologists". ARRT. Retrieved 2014-01-05. 
  5. "Individual State Licensure Information". American Society of Radiologic Technologists. Retrieved 19 July 2013. 
  6. 6.0 6.1 Smith-Bindman R, Lipson J, Marcus R, et al. (December 2009). "Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer". Arch. Intern. Med. 169 (22): 2078–86. doi:10.1001/archinternmed.2009.427. PMID 20008690. 
  7. 7.0 7.1 Berrington de González A, Mahesh M, Kim KP, et al. (December 2009). "Projected cancer risks from computed tomographic scans performed in the United States in 2007". Arch. Intern. Med. 169 (22): 2071–7. doi:10.1001/archinternmed.2009.440. PMID 20008689. 
  8. 8.0 8.1 8.2 8.3 8.4 8.5 8.6 8.7 Brenner DJ, Hall EJ (November 2007). "Computed tomography – an increasing source of radiation exposure". N. Engl. J. Med. 357 (22): 2277–84. doi:10.1056/NEJMra072149. PMID 18046031. 
  9. Tubiana M (February 2008). "Comment on Computed Tomography and Radiation Exposure". N. Engl. J. Med. 358 (8): 852–3. doi:10.1056/NEJMc073513. PMID 18287609. 
  10. CT Screening
  11. 11.0 11.1 11.2 American Headache Society (September 2013), "Five Things Physicians and Patients Should Question", Choosing Wisely: an initiative of the ABIM Foundation (American Headache Society), retrieved 10 December 2013 , which cites
  12. Wiener, R. S.; Gould, M. K.; Woloshin, S.; Schwartz, L. M.; Clark, J. A. (2012). ""What do you mean, a spot?": A qualitative analysis of patients' reactions to discussions with their doctors about pulmonary nodules". CHEST Journal 143 (3): 672–677. doi:10.1378/chest.12-1095. PMC 3590883. PMID 22814873. 
  13. 13.0 13.1 13.2 American College of Chest Physicians; American Thoracic Society (September 2013), "Five Things Physicians and Patients Should Question", Choosing Wisely: an initiative of the ABIM Foundation (American College of Chest Physicians and American Thoracic Society), retrieved 6 January 2013 , which cites
  14. "Ankle Fractures". orthoinfo.aaos.org. American Association of Orthopedic Surgeons. Retrieved 2010-05-30. 
  15. Buckwalter, Kenneth A. et al. (11 September 2000). "Musculoskeletal Imaging with Multislice CT". ajronline.org. American Journal of Roentgenology. Retrieved 2010-05-22. 
  16. Brian R. Subach M.D., F.A.C.S et al."Reliability and accuracy of fine-cut computed tomography scans to determine the status of anterior interbody fusions with metallic cages"
  17. 17.0 17.1 17.2 17.3 17.4 17.5 Redberg, Rita F., and Smith-Bindman, Rebecca. "We Are Giving Ourselves Cancer", New York Times, Jan. 30, 2014
  18. 18.0 18.1 18.2 18.3 18.4 Semelka, RC; Armao, DM; Elias, J, Jr.; Huda, W. (May 2007). "Imaging strategies to reduce the risk of radiation in CT studies, including selective substitution with MRI.". J Magn Reson Imaging 25 (5): 900–9. doi:10.1002/jmri.20895. PMID 17457809. 
  19. 19.0 19.1 19.2 19.3 19.4 19.5 19.6 19.7 19.8 19.9 19.10 19.11 19.12 19.13 19.14 19.15 Furlow, B (May–Jun 2010). "Radiation dose in computed tomography.". Radiologic technology 81 (5): 437–50. PMID 20445138. 
  20. 20.0 20.1 20.2 20.3 20.4 Davies, H. E.; Wathen, C. G.; Gleeson, F. V. (25 February 2011). "The risks of radiation exposure related to diagnostic imaging and how to minimise them". BMJ 342 (feb25 1): d947–d947. doi:10.1136/bmj.d947. 
  21. Baysson, H; Etard, C; Brisse, HJ; Bernier, MO (January 2012). "[Diagnostic radiation exposure in children and cancer risk: current knowledge and perspectives].". Archives de pediatrie : organe officiel de la Societe francaise de pediatrie 19 (1): 64–73. doi:10.1016/j.arcped.2011.10.023. PMID 22130615. 
  22. Larson DB, Rader SB, Forman HP, Fenton LZ (August 2007). "Informing parents about CT radiation exposure in children: it's OK to tell them". AJR Am J Roentgenol 189 (2): 271–5. doi:10.2214/AJR.07.2248. PMID 17646450. 
  23. 23.0 23.1 Namasivayam, S; Kalra, MK; Torres, WE; Small, WC (Jul 2006). "Adverse reactions to intravenous iodinated contrast media: a primer for radiologists.". Emergency radiology 12 (5): 210–5. doi:10.1007/s10140-006-0488-6. PMID 16688432. 
  24. 24.0 24.1 24.2 Christiansen, C (2005-04-15). "X-ray contrast media – an overview.". Toxicology 209 (2): 185–7. doi:10.1016/j.tox.2004.12.020. PMID 15767033. 
  25. 25.0 25.1 Wang H, Wang HS, Liu ZP (October 2011). "Agents that induce pseudo-allergic reaction". Drug Discov Ther 5 (5): 211–9. PMID 22466368. 
  26. Drain, KL; Volcheck, GW (2001). "Preventing and managing drug-induced anaphylaxis.". Drug safety : an international journal of medical toxicology and drug experience 24 (11): 843–53. doi:10.2165/00002018-200124110-00005. PMID 11665871. 
  27. editor, Mariana C. Castells,. Anaphylaxis and hypersensitivity reactions. New York: Humana Press. p. 187. ISBN 9781603279505. 
  28. 28.0 28.1 Hasebroock, KM; Serkova, NJ (April 2009). "Toxicity of MRI and CT contrast agents.". Expert opinion on drug metabolism & toxicology 5 (4): 403–16. doi:10.1517/17425250902873796. PMID 19368492. 
  29. 29.0 29.1 29.2 Cuttler JM, Pollycove M (2009). "Nuclear energy and health: and the benefits of low-dose radiation hormesis". Dose Response 7 (1): 52–89. doi:10.2203/dose-response.08-024.Cuttler. PMC 2664640. PMID 19343116. 
  30. 30.0 30.1 "What are the Radiation Risks from CT?". Food and Drug Administration. 2009. 
  31. 31.0 31.1 31.2 31.3 31.4 31.5 Hall, EJ; Brenner, DJ (May 2008). "Cancer risks from diagnostic radiology.". The British journal of radiology 81 (965): 362–78. doi:10.1259/bjr/01948454. PMID 18440940. 
  32. 32.0 32.1 32.2 32.3 32.4 Shrimpton, P.C; Miller, H.C; Lewis, M.A; Dunn, M. Doses from Computed Tomography (CT) examinations in the UK – 2003 Review
  33. Poston, edited by Michael T. Ryan, John W. (2005). A half century of health physics. Baltimore, Md.: Lippincott Williams & Wilkins. p. 164. ISBN 9780781769341. 
  34. Polo SE, Jackson SP (March 2011). "Dynamics of DNA damage response proteins at DNA breaks: a focus on protein modifications". Genes Dev. 25 (5): 409–33. doi:10.1101/gad.2021311. PMC 3049283. PMID 21363960. 
  35. The Measurement, Reporting, and Management of Radiation Dose in CT "It is a single dose parameter that reflects the risk of a nonuniform exposure in terms of an equivalent whole-body exposure."
  36. Wintermark M, Lev MH (January 2010). "FDA investigates the safety of brain perfusion CT". AJNR Am J Neuroradiol 31 (1): 2–3. doi:10.3174/ajnr.A1967. PMID 19892810. 
  37. "Image Gently". The Alliance for Radiation Safety in Pediatric Imaging. Retrieved 19 July 2013. 
  38. "Image Wisely". Joint Task Force on Adult Radiation Protection. Retrieved 19 July 2013. 
  39. "Optimal levels of radiation for patients". World Health Organization. Retrieved 19 July 2013. 
  40. "Global Initiative on Radiation Safety in Healthcare Settings". World Health Organization. Retrieved 19 July 2013. 
  41. Andrew Skelly (Aug 3, 2010). "CT ordering all over the map". The Medical Post. 
  42. Korley FK, Pham JC, Kirsch TD (October 2010). "Use of advanced radiology during visits to US emergency departments for injury-related conditions, 1998–2007". JAMA 304 (13): 1465–71. doi:10.1001/jama.2010.1408. PMID 20924012. 
  43. Contrast agent for radiotherapy CT (computed tomography) scans. Patient Information Series No. 11 at University College London Hospitals NHS Foundation Trust. Last reviewed: October 2009
  44. Dahlman, P.; Semenas, E.; Brekkan, E.; Bergman, A.; Magnusson, A. (2000). "Detection and Characterisation of Renal Lesions by Multiphasic Helical Ct". Acta Radiologica 41 (4): 361–366. doi:10.1080/028418500127345479. PMID 10937759. 
  45. Computerized Tomography chapter at University of Connecticut Health Center.
  46. Udupa, J.K. and Herman, G. T., 3D Imaging in Medicine, 2nd Edition, CRC Press, 2000
  47. Bhowmik, Ujjal Kumar; Zafar Iqbal, M.; Adhami, Reza R. (28 May 2012). "Mitigating motion artifacts in FDK based 3D Cone-beam Brain Imaging System using markers". Central European Journal of Engineering 2 (3): 369–382. Bibcode:2012CEJE....2..369B. doi:10.2478/s13531-012-0011-7. 
  48. Boas FE and Fleischmann D (2011). "Evaluation of Two Iterative Techniques for Reducing Metal Artifacts in Computed Tomography". Radiology 259 (3): 894–902. doi:10.1148/radiol.11101782. PMID 21357521. 
  49. Mouton, A., Megherbi, N., Van Slambrouck, K., Nuyts, J., Breckon, T.P. (2013). "An Experimental Survey of Metal Artefact Reduction in Computed Tomography". Journal of X-Ray Science and Technology. doi:10.3233/XST-130372. 
  50. J. Nemirovsky, A. Lifshitz, and I. Be’erya (May 2011). "Tomographic reconstruction of incompressible flow". AIP Rev. Sci. Instrum. 82 (5). Bibcode:2011RScI...82e5115N. doi:10.1063/1.3590934. 
  51. Van de Casteele E and Van Dyck D and Sijbers J and Raman E (2004). "A model-based correction method for beam hardening artefacts in X-ray microtomography". Journal of X-ray Science and Technology 12 (1): 43–57. 
  52. Van Gompel G and Van Slambrouck K and Defrise M and Batenburg KJ and Sijbers J and Nuyts J (2011). "Iterative correction of beam hardening artifacts in CT". Medical Physics 38 (1): 36–49. doi:10.1118/1.3577758. 
  53. Simpson, Graham (2009). "Thoracic computed tomography: principles and practice" (PDF). Australian Prescriber, 32:4. Retrieved September 25, 2009.
  54. "Imaging Techniques in Conservation". Journal of Conservation and Museum Studies (Ubiquity Press): 17–29. 2012. doi:10.5334/jcms.1021201. Retrieved 17 March 2013.  Unknown parameter |autho r= ignored (help)
  55. Megherbi, N., Flitton, G.T., Breckon, T.P. (September 2010). "A Classifier based Approach for the Detection of Potential Threats in CT based Baggage Screening". Proc. International Conference on Image Processing. IEEE. pp. 1833–1836. doi:10.1109/ICIP.2010.5653676. Retrieved 5 November 2013. 
  56. Megherbi, N., Han, J., Flitton, G.T., Breckon, T.P. (September 2012). "A Comparison of Classification Approaches for Threat Detection in CT based Baggage Screening". Proc. International Conference on Image Processing. IEEE. pp. 3109–3112. doi:10.1109/ICIP.2012.6467558. Retrieved 5 November 2013. 
  57. Flitton, G.T., Breckon, T.P., Megherbi, N. (September 2013). "A Comparison of 3D Interest Point Descriptors with Application to Airport Baggage Object Detection in Complex CT Imagery". Pattern Recognition (Elsevier) 46 (9): 2420–2436. doi:10.1016/j.patcog.2013.02.008. Retrieved 5 November 2013. 
  58. 58.0 58.1 Littleton, J.T. "Conventional Tomography". A History of the Radiological Sciences. American Roentgen Ray Society. Retrieved 11 January 2014. 
  59. Radon J (1917). "Uber die Bestimmung von Funktionen durch ihre Integralwerte Langs Gewisser Mannigfaltigkeiten (English translation: On the determination of functions from their integrals along certain manifolds)". Ber. Saechsische Akad. Wiss. 29: 262. 
  60. Radon, Johann (1 December 1986). "On the determination of functions from their integral values along certain manifolds". IEEE Transactions on Medical Imaging 5 (4): 170–176. doi:10.1109/TMI.1986.4307775. PMID 18244009. 
  61. Hornich H., Translated by Parks PC. A Tribute to Johann Radon. IEEE Trans. Med. Imaging. 1986;5(4) 169–9.
  62. Kaczmarz S (1937). "Angenäherte Auflösung von Systemen linearer Gleichungen". Bulletin International de l'Académie Polonaise des Sciences et des Lettres. Classe des Sciences Mathématiques et Naturelles. Série A, Sciences Mathématiques. 35: 355–7. 
  63. Kaczmarz S., "Approximate solution of system of linear equations. Int. J. Control. 1993; 57-9.
  64. Bracewell RN., [Strip Integration in Radio Astronomyhttp://articles.adsabs.harvard.edu//full/1956AuJPh...9..198B/0000198.000.html]. Aust. J. Phys. 1956;9: 198–217.
  65. Oldendorf WH. Isolated flying spot detection of radiodensity discontinuities – displaying the internal structural pattern of a complex object. Ire Trans Biomed Electron. 1961 Jan;BME-8:68–72.
  66. Oldendorf WH (Jun 1978). "The quest for an image of brain: a brief historical and technical review of brain imaging techniques". Neurology 28 (6): 517–33. 
  67. F. Natterer, "The Mathematics of Computerized Tomography (Classics in Applied Mathematics)", Society for Industrial Mathematics, isbn= 0898714931
  68. F. Natterer and F. Wübbeling "Mathematical Methods in Image Reconstruction (Monographs on Mathematical Modeling and Computation)", Society for Industrial (2001), isbn= 0898714729
  69. Deuflhard, P.; Dössel, O.; Louis, A. K.; Zachow, S. (5 March 2009). "More Mathematics into Medicine!". Zuse Institute Berlin. p. 2. 
  70. Allen M.Cormack: My Connection with the Radon Transform, in: 75 Years of Radon Transform, S. Gindikin and P. Michor, eds., International Press Incorporated (1994), pp. 32–35, ISBN 1-57146-008-X
  71. Richmond, Caroline (September 18, 2004). "Obituary – Sir Godfrey Hounsfield". BMJ (London, UK: BMJ Group). 2004:329:687 (18 September 2004). Retrieved September 12, 2008. 
  72. 72.0 72.1 BECKMANN, E. C. (January 2006). "CT scanning the early days". TheBritish Journal of Radiology 79 (937): 5–8. doi:10.1259/bjr/29444122. PMID 16421398. 
  73. "The Beatles greatest gift... is to science". Whittington Hospital NHS Trust. Retrieved 2007-05-07. 
  74. Maizlin, Zeev V.; Vos, Patrick M. (2012). "Do we really need to thank the Beatles for the financing of the development of the computed tomography scanner?". Journal of Computer Assisted Tomography 36 (2): 161–164. doi:10.1097/RCT.0b013e318249416f. PMID 22446352. 
  75. "The Nobel Prize in Physiology or Medicine 1979 Allan M. Cormack, Godfrey N. Hounsfield". Nobelprize.org. Retrieved 19 July 2013. 
  76. Filler, Aaron G. (2010). "The History, Development and Impact of Computed Imaging in Neurological Diagnosis and Neurosurgery: CT, MRI, and DTI". The Internet Journal of Neurosurgery 7 (1). doi:10.5580/23c6. 
  77. "This Machine Can 3-D Scan Your Insides In A Single Heartbeat". Fastcoexist.com. January 3, 2014. Retrieved 2014-01-05. 
  78. Tomography, X-Ray Computed at the US National Library of Medicine Medical Subject Headings (MeSH)
  79. Retsky, Michael (31 July 2008). "Electron beam computed tomography: Challenges and opportunities". Physics Procedia 1 (1): 149–154. doi:10.1016/j.phpro.2008.07.090. 
  80. Novelline, Robert. Squire's Fundamentals of Radiology. Harvard University Press. 5th edition. 1997. ISBN 0-674-83339-2.

External links

This article is issued from Wikipedia. The text is available under the Creative Commons Attribution/Share Alike; additional terms may apply for the media files.